27
Pump Too Hot
This alarm is activated when the temperature
inside the pump exceeds 55°C.
“Pump too hot!” is displayed at the top of the
“standard” Run screen (screen A) and an
audible alarm sounds. Therapy can continue.
Make sure the pump is connected to mains
power if this alarm is activated, and check that
the ventilation holes on the pump rear case
are not covered or blocked.
When the temperature inside the pump
exceeds 60°C, the full alarm is activated.
Screen B is permanently displayed and an
audible alarm sounds. Therapy is suspended.
Switch off the pump and allow pump to cool.
Then switch back on and continue therapy.
If either alarm continues, switch off the
pump and call the service engineer.
Condition
Description and Corrective Action
A
B
Pump too hot!
Stop