16
1. The visual-only warning is a fault message on the
LCD screen. The LED indications remain
unchanged and there is no audible alarm.
2. If the fault is not cleared then the warning is
replaced by an audible and visual alarm, which
consists of:
• A fault message on the LCD screen.
• The LED indicators on the front of the pump
change to red and start flashing.
• An audible alarm will sound which increases in
pitch and intensity if the alarm is ignored.
The warning and alarm can be cleared by either:
• Checking and rectifying the fault on the system, or
• Pressing the button below “Stop” to put the pump
into Standby.
Refer to Section 7, Page 22 “Troubleshooting” for the
alarms, their possible causes and their remedies.
Changing Options
To change the pump operating options, the pump must
be in Standby. If the pump is in Run mode, press and
hold the button below “Stop” for 2 seconds to put the
pump into Standby.
On the Standby screen, press the button below
“Options” and the “Options” menu is displayed:
• The “border” around an item indicates the currently
selected item.
• To select a new menu item, press the buttons below
the “up” and “down” arrows.
• To confirm the new menu item, press the button
below the “tick”.
• To exit and return to the Standby screen, select and
confirm “Exit” on the menu.
1. To change the Audio Alarm Volume setting: