Error codes
Error codes are shown on the display. They are used to
indicate a problem with the weighing system; this may
be due to an operator error or a possible fault condition.
The table below shows common error codes with their
meanings and possible causes.
Display
Condition
Probable cause
Zero error
User attempted to set a zero point
more than 77 lbs. or 35 kg above the
factory calibrated zero
Overload error
The total load on the bed exceeds 559
lbs. or 254.5 kg
Underload error
The total load on the bed is more than
22 lbs. or 10 kg below the zero point
Auto Compensation
negative error
Auto Compensation was attempted for
a weight decrease greater than 220
lbs. or 100 kg.
Auto Compensation
positive error
Auto Compensation was attempted for
a weight increase greater than 220 lbs.
or 100 kg.
Unstable weight
reading
A stable weight reading was not
obtained within 10 seconds of pressing
the
Weigh
key
If the display shows an error code other than those given
above, contact the ArjoHuntleigh Service Department.
41
746-505-AM-6