Attendant Control Panel additional features
Function
lockout
Under certain circumstances it may be necessary to
prevent operation of the bed controls, e.g. when traction
equipment is being used.
To selectively disable bed functions:
Press the
Function Lockout
key on the Attendant Control
Panel. The indicator above the key will light.
Press the Attendant Control Panel function key
corresponding to the function to be disabled. The indicator
above the key will light yellow.
To disable ALL bed functions simultaneously:
Press and hold the
Function Lockout
key for five
seconds.
If no function key is pressed within five seconds of pressing the
Function Lockout
key, the indicator above the key will go out
and the Lockout function must be selected again.
Function Lockout settings are retained when the bed is
disconnected from the electrical supply.
To reinstate previously disabled bed functions:
Press the
Function Lockout
key on the Attendant Control
Panel. The indicator above the key will light.
Press the Attendant Control Panel function key
corresponding to the function to be reinstated. The
indicator above the key will go out.
29
746-505-AM-6