001-1
1900-EN rev
. 5
Care and Maintenance
12
Troubleshooting
Preventive Maintenance Schedule
The
Semi-Permanent Rack
is subject to wear and the following actions must be performed when specified
to ensure that the product remains within its original manufacturing specification.
WARNING
: The points on this checklist are the minimum the manufacturer recommends. In some cases
more frequent inspections should be carried out. Continuing to use the
Semi-Permanent Rack
without
conducting regular inspections will seriously compromise the user's and patient’s safety. Preventive
maintenance specified in this manual can prevent accidents.
WARNING:
Safety related maintenance and authorized service must be carried out by qualified
personnel, fully trained in servicing procedures by Arjo and equipped with correct tools and proper
documentation, including the parts list and service manual. Failure to meet these requirements could
result in personal injuries and / or unsafe equipment.
POINTS TO BE INSPECTED BY
USER / SERVICE TECHNICIAN
FREQUENCY
Annually or 1400 cycles
Monthly
Before every use
Inspect for evidence of external damage or missing parts.
Make sure all bolts and nuts are properly tightened.
Make sure that bases are properly adjusted.
Make sure track is properly assembled to the posts.
Make sure that all posts are straight.
Clean the inside of the tracks to ensure there is no dust or grease build-up inside the tracks.
Check to make sure that the trolleys move freely along the tracks.
WARNING
: If any of the above inspection point fails, DO NOT USE the
Semi-Permanent Rack
until
appropriate corrective action is taken.
Annual Inspection by an Authorized Technician
The
Semi-Permanent Rack
must be inspected by a qualified Arjo technician on an annual basis to ensure
the proper functioning of its components, as well as the safety of the patients and care givers that use the
Semi-Permanent Rack.
Please contact your local representative to make arrangements.
NOTE
: If the product does not work as intended, immediately contact your local Arjo agent for support.
Cleaning
To clean the
Semi-Permanent Rack
, wipe it down with a damp cloth using warm water and mild detergent.
CAUTION
: Do not use phenol, chlorine or any other type of solvent that may damage the finish.
To ensure a better rolling surface for the trolley wheels, clean the inside of the track every four months. To
do so, insert a damp cloth in the opening and slide it from one end of the track to the other.
Care and Maintenance