6. Decontamination
The following processes are recommended, but should be adapted to comply with the local or
national guidelines (Decontamination of Medical Devices) which may apply within the Healthcare
Facility or the country of use. If you are uncertain, you should seek advice from your local Infection
Control Specialist.
The
Alpha Trancell
Deluxe system should be routinely decontaminated between patients and at
regular intervals while in use; as is good practice for all reusable medical devices.
To Clean
Clean all exposed surfaces and remove any organic debris by wiping
with a cloth moistened with a simple (neutral) detergent and water.
Dry thoroughly.
Do not allow water or cleaning solutions to collect on the surface of
the pump.
Chemical Disinfection
To protect the integrity of the cover we recommend a chlorine-
releasing agent, such as sodium hypochlorite, at a strength of
1,000ppm available chlorine (this may vary from 250ppm to
10,000ppm depending on local policy and contamination status).
Wipe all cleaned surfaces with the solution, then wipe with a cloth
moistened in water and dry thoroughly.
Alcohol based disinfectants (strength 70%) may be used as an
alternative.
Ensure the product is dry before storage.
If an alternative disinfectant is selected from the wide variety
available we recommend that suitability for use is confirmed with the
chemical supplier prior to use.
DO NOT WRING/MANGLE, AUTOCLAVE OR USE PHENOLIC
BASED SOLUTIONS.
Thermal Disinfection
For mattress top cover information including laundering guidelines,
refer to “Cover Specification” on page 19.
Re-use with multiple
patients
Professional hygiene maintenance is required before re-use of the
system with a different patient.
WARNING
Remove the electrical supply to the pump by disconnecting the mains power cord
from the mains power supply before cleaning.
Protective clothing should always be worn when carrying out decontamination
procedures.
Caution
Do not use Phenol-based solutions or abrasive compounds or pads during the
decontamination process as these will damage the surface coating. Do not boil or
autoclave the cover.
Avoid immersing electrical parts in water during the cleaning process. Do not
spray cleaning solutions directly onto the pump.