7. Routine Maintenance
Alpha Trancell
Deluxe Systems
Maintenance
The equipment has been designed to be virtually maintenance-free
between service periods.
Servicing
Arjo will make available on request service manuals, component
parts lists and other information necessary for Arjo trained
personnel to repair the system
Service Period
Arjo recommend that the
Alpha Trancell
Deluxe systems should be
serviced after 12 months continuous running time, by an Arjo
authorised service agent.
Alpha Trancell
Deluxe Pump
General Care,
Maintenance and
Inspection
Check all electrical connections and the mains power cable for signs
of wear or damage.
If the pump has been subjected to abnormal treatment, e.g.
immersed in water or dropped, the unit must be returned to an
authorised service centre.
Mattresses and Seats
General Care
Remove the cover from the mattress or seat.
Inspect the cover for signs of wear or any tears, and check that all
cover fasteners are secure.
Check the security of all internal connections, including:
• Between the cells and the manifold.
• To the rapid deflate control.
Make sure all cell fasteners are correctly connected to the mattress
base sheet and are not loose or damaged.
Serial Number Labels
Pump
The serial number label is on the back of the pump case.
Mattress and Seat
The serial number label is on the base cover.
Quote these serial numbers when requesting service.