17
9. Technical Specification
Pump
Model:
Alpha Trancell
Deluxe
USA
Supply Voltage:
120 V
Supply Frequency:
60 Hz
Power Input:
14 VA
Size:
248 x 160 x 116 mm (9.8 x 6.3 x 4.6”)
Weight:
2.7 kg (6lb)
Case Material:
ABS Plastic
Plug Fuse Rating:
5A to BS1362 (UK ONLY)
Pump Fuse Rating:
2 x T1AL 250 V
Degree of protection
against electric shock:
Mains Connected - Class II, Double Insulated with functional earth Type
BF
Degree of protection
against liquid ingress:
IP21 - Protection against ingress of solid objects more than 12.5mm
diameter and water droplets falling vertically.
Mode of operation:
Continuous
Environmental Information
Condition
Temperature Range
Relative Humidity
Atmospheric Pressure
Operating
+10°C to +40°C
(+50°F to +104°F)
30% to 75%
(non-condensing)
700hPa to 1060 hPa
Storage (Long Term)
+10°C to +40°C
(+50°F to +104°F)
20% to 95%
(non-condensing)
700 hPa to 1060 hPa
Storage (Short Term)
-20°C to +65°C
(-4°F to 149°F)
20% to 95%
(non-condensing)
500 hPa to 1060 hPa
NOTE
If the pump is stored in conditions outside the “Operating” ranges, allow time for its temperature
to stabilise to normal before use.
NOTE
One of the effects of prolonged exposure to high temperatures is to increase the self-discharge
of the internal battery; this will reduce the duration of power fail alarms. The pump will fully charge
the battery over a 24-hour period when the pump is connected to a mains power supply.