PUMP CONFIGURATIONS
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
7 / 48
7. After entering an address e.g. 5 press the
key “MODE”. The flashing decimal points
will change to the right hand display.
9. Press key “START” to ac-
knowledge the value en-
tered. The flashing decimal
points change back to the
left hand display.
8. Enter now the value on the right hand display. The
range of the value is given by the table in
chap.2.6
10. To quit the configuration mode press the
“ON/OFF” key at least 2 seconds.
Changes in configuration become active,
after the pump is switched on normally again.
2. Press key “START”
3. Press key “MODE”
4. Press key “START”
again. The display will
show “Cod ----“.
2.3.3. Step 3
Important remark:
Invalid values entered will be corrected automatically by the pump to the maxima or
minima value allowed for the according address!
2.4. First activation of a PIN code (write protection)
The activation of a PIN code allows you to protect the configuration from unauthor-
ized access. To activate the PIN code,
enter into the configuration mode.
5
...
.
1.
To enter into the configuration mode, switch the pump on
while keeping the keys “MODE” and “START/STOP” pressed.
707
5. Press “MODE” key (Code “0” will be set) then “START”
key to acknowledge the entered PIN code “0”.
Ro
P
g
0