REPLACEMENT OF PARTS
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
39 / 48
1. Remove the four screws on the back side.
2. Separate the
front and the
back side.
3. Disconnect the battery, the
supply and the docking in-
terface connection.
7. REPLACEMENT
OF
PARTS
7.1. General
CAUTION!
The ARGUS 707 V may only be used with accessories and spare parts which have
been approved by ARGUS Medical AG for safe technical use.
CAUTION!
If a door, a housing, a pressure sensor or a main board is replaced, a full
calibration (pressure sensors and volume calibration) is required.
7.2. Disassembling of the ARGUS 707 V
CAUTION!
Disconnect the mains cable from the power outlet before opening the housing!
Observe the antistatic protection rules when disassembling the ARGUS 707 V
(the use of an antistatic table mat and a grounded clip are recommended).
7.2.1. Disassembly of the case