background image

 

PUMP CONFIGURATIONS 

14.203.B_A707 en.SM.V4.3X 

ARGUS Medical AG  28.03.06 / PJ 

17 / 48 

720 

No 

Phentolamine 

No / Yes 

 

761 

No 

Urokinase 

No / Yes 

721 

No 

Phenylephrin 

No / Yes 

 

762 

No 

Vasopressine 

No / Yes 

722 

No 

Pirarubicine 

No / Yes 

 

763 

No 

Vecuronium 

No / Yes 

723 

No 

Plicamycin 

No / Yes 

 

764 

No 

Verapamil 

No / Yes 

724 

No 

Procainamide 

No / Yes 

 

765 

No 

Vinblastine 

No / Yes 

725 

No 

Procarbazine 

No / Yes 

 

766 

No 

Vincristine 

No / Yes 

726 

No 

Propafenon 

No / Yes 

 

767 

No 

Vindesine 

No / Yes 

727 

No 

Propofol 

No / Yes 

 

768 

No 

Vinorelbine 

No / Yes 

728 

No 

Prostaglandin e1 

No / Yes 

 

769 

No 

User defined med.  1 

No / Yes 

729 

No 

Raltitrexed 

No / Yes 

 

770 

No 

User defined med.  2 

No / Yes 

730 

No 

Rapilysin 

No / Yes 

 

771 

No 

User defined med.  3 

No / Yes 

731 

No 

Remifentanyl 

No / Yes 

 

772 

No 

User defined med.  4 

No / Yes 

732 

No 

Risordan 

No / Yes 

 

773 

No 

User defined med.  5 

No / Yes 

733 

No 

Ritodrine hcl 

No / Yes 

 

774 

No 

User defined med.  6 

No / Yes 

734 

No 

Rituximab 

No / Yes 

 

775 

No 

User defined med.  7 

No / Yes 

735 

No 

Rocuronium bromide 

No / Yes 

 

776 

No 

User defined med.  8 

No / Yes 

736 

No 

Ropivacane 

No / Yes 

 

777 

No 

User defined med.  9 

No / Yes 

737 

No 

Salbutamol 

No / Yes 

 

778 

No 

User defined med. 10 

No / Yes 

738 

No 

Somatostatin 

No / Yes 

 

779 

No 

User defined med. 11 

No / Yes 

739 

No 

Streptokinase 36000 ie 

No / Yes 

 

780 

No 

User defined med. 12 

No / Yes 

740 

No 

Streptokinase 1000 ie 

No / Yes 

 

781 

No 

User defined med. 13 

No / Yes 

741 

No 

Streptokinase 3000 ie 

No / Yes 

 

782 

No 

User defined med. 14 

No / Yes 

742 

No 

Streptokinase 6000 ie 

No / Yes 

 

783 

No 

User defined med. 15 

No / Yes 

743 

No 

Streptokinase 

No / Yes 

 

784 

No 

User defined med. 16 

No / Yes 

744 

No 

Streptozocine 

No / Yes 

 

785 

No 

User defined med. 17 

No / Yes 

745 

No 

Succinylcholine chloride 

No / Yes 

 

786 

No 

User defined med. 18 

No / Yes 

746 

No 

Sufentanil 

No / Yes 

 

787 

No 

User defined med. 19 

No / Yes 

747 

No 

T.P.N 

No / Yes 

 

788 

No 

User defined med. 20 

No / Yes 

748 

No 

Tegafur 

No / Yes 

 

789 

No 

User defined med. 21 

No / Yes 

749 

No 

Temozolomide 

No / Yes 

 

790 

No 

User defined med. 22 

No / Yes 

750 

No 

Terbutaline 

No / Yes 

 

791 

No 

User defined med. 23 

No / Yes 

751 

No 

Theopyllin 

No / Yes 

 

792 

No 

User defined med. 24 

No / Yes 

752 

No 

Thiotepa 

No / Yes 

 

793 

No 

User defined med. 25 

No / Yes 

753 

No 

Tirofiban 

No / Yes 

 

794 

No 

User defined med. 26 

No / Yes 

754 

No 

Tolazolin hcl 

No / Yes 

 

795 

No 

User defined med. 27 

No / Yes 

755 

No 

Topotecan 

No / Yes 

 

796 

No 

User defined med. 28 

No / Yes 

756 

No 

Trastuzumab 

No / Yes 

 

797 

No 

User defined med. 29 

No / Yes 

757 

No 

Trimethaphan camsylate 

No / Yes 

 

798 

No 

User defined med. 30 

No / Yes 

758 

No 

Umbilical arterial catheter 

No / Yes 

 

799 

No 

User defined med. 31 

No / Yes 

759 

No 

Umbilical venous catheter 

No / Yes 

 

800 

No 

User defined med. 32 

No / Yes 

760 

No 

Urapidil 

No / Yes 

 

 

 

 

 

 
 
 

REMARK : 

Via barcode reader all medication names can be selected, even if they are not released in 
the configuration. 
 

 
 
 
 
 
 
 
 

Summary of Contents for 707 V

Page 1: ...3 B_A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 1 48 Service Manual for Volumetric Infusion Pump ARGUS 707 V Made in Switzerland ARGUS Medical AG CH 3627 Heimberg Switzerland a member of the CODAN g...

Page 2: ...communication of the pump 18 3 1 General 18 3 2 Serial communication protocol 18 4 ARGUS service 19 4 1 General 19 4 2 ARGUS service Configuration 20 4 3 ARGUS service Calibration volumetric pums only...

Page 3: ...707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 2 48 Global alarm Display Drop detector Drop chamber Park area for drop detector Air bubble detector Pressure sensor downstream automatic free flow clamp P...

Page 4: ...M V4 3X ARGUS Medical AG 28 03 06 PJ 3 48 AC power connection DC connection QUICK 100 Docking Station interface Staff alert and RS 232 connection Drop detector connection Identification plate with pum...

Page 5: ...ARGUS 707 V infusion pump may only be used with spare parts accessories consumables and IV sets with Luer Lock connections recommended by ARGUS Medical AG The functional safety of the pump is not guar...

Page 6: ...without the possibility to modify any configurations For a complete overview please take the ARGUS service PC tool Flashing decimal points indicate which display is ready to accept an input by the ke...

Page 7: ...ng the keypad Please refer to chapter 2 6 where the meaning of the adresses are explained To have access to all configuration options please use the ARGUS service PC tool 2 3 1 Step 1 2 3 2 Step 2 You...

Page 8: ...ecome active after the pump is switched on normally again 2 Press key START 3 Press key MODE 4 Press key START again The display will show Cod 2 3 3 Step 3 Important remark Invalid values entered will...

Page 9: ...t PIN code e g Please note The interrogation mode can always be accessed without the PIN 2 5 Changing an existing PIN code Enter the configuration mode using present PIN select add 0 and set the new c...

Page 10: ...tion without drop detector 0 No 1 Yes 3 11 No Recall of the last used ml h rate 0 No 1 Yes 4 19 No Buzzer at start 0 No 1 Yes 5 44 Yes Automatic pressure release 0 No 1 Yes 6 49 No Alarm acknowledge w...

Page 11: ...10 No trA No Yes 19 4 No Buzzer at start No Yes 20 No Menu Clr clear ml inf 15 No No Yes 21 22 No Menu trA transport 10 18 No No Yes 23 Yes Menu PrL pressure alarm limit No Yes 24 Yes Menu CAP battery...

Page 12: ...1 976 Correction at 250 ml h ratio 900 1100 212 967 Correction at 300 ml h ratio 900 1100 213 952 Correction at 400 ml h ratio 900 1100 214 942 Correction at 500 ml h ratio 900 1100 215 925 Correction...

Page 13: ...262 1054 Correction at 3 ml h ratio 900 1100 263 1044 Correction at 5 ml h ratio 900 1100 264 1035 Correction at 10 ml h ratio 900 1100 265 1027 Correction at 25 ml h ratio 900 1100 266 1020 Correctio...

Page 14: ...ml h 1 999 342 1200 Max bolus rate ml h 1 1200 343 10 Max bolus total ml 1 999 344 10 Upstream occlusion sensitivity factor 0 60 345 20 Air summation volume 50 l 2 40 346 7 Downstream default pressure...

Page 15: ...ASCII Using the ARGUS service tool the complete and detailed pump configuration can be done 2 8 Special configuration options with ARGUS service Enabling a new IV set Follow the steps below to enable...

Page 16: ...serM 7 User med name 7 char 1 8 ASCII 568 UserM 8 User med name 8 char 1 8 ASCII 569 UserM 9 User med name 9 char 1 8 ASCII 570 UserM 10 User med name 10 char 1 8 ASCII 571 UserM 11 User med name 11 c...

Page 17: ...stine No Yes 687 No Mannitol No Yes 628 No Cefazolin sodium No Yes 688 No Metaraminol bitartrate No Yes 629 No Central line No Yes 689 No Methohexital sodium No Yes 630 No Ceruletid No Yes 690 No Meth...

Page 18: ...o Yes 740 No Streptokinase 1000 ie No Yes 781 No User defined med 13 No Yes 741 No Streptokinase 3000 ie No Yes 782 No User defined med 14 No Yes 742 No Streptokinase 6000 ie No Yes 783 No User define...

Page 19: ...or ARGUS 100 M the software switches automatically to the docking interface and the docking station builds the separation device galvanic isolation then 3 2 Serial communication protocol The followin...

Page 20: ...pump and the ARGUS docking station This software may be available from your local distribu tor or directly from ARGUS REMARK ARGUS service may only be used with software versions greater or equal to...

Page 21: ...tioned in chapter 2 7 can be modified herein in its own tree structure as shown below All indexes which are different from the pump firmware default are high lighted Calibration part 2 Details of the...

Page 22: ...the according index 6 Modify the according index within the given restrictions shown Each value number must be acknowledged by the green Enter button Go through point 5 6 for all further indexes you w...

Page 23: ...n factor 4 4 ARGUS service Toolbox With the ARGUS service PC tool you can set the pump clock change PIN code read and print out history etc Select the next step by pressing one of the but tons set clo...

Page 24: ...oustic sound The point in time when an active reminder alarm occurs is given by the settings of the configuration 393 and 394 and the pump internal clock Any value higher than 0 on those indexes will...

Page 25: ...p mode Airbubble pump stop Transport off Enter PC configuration mode Door open pump stop Transport on Exit setup or PC configuration mode No drops pump stop Data lock off Downstream occlusion pump sto...

Page 26: ...ARGUS service 14 203 B_A707 en SM V4 3X ARGUS Medical AG 28 03 06 PJ 25 48 All pre alarms alarms and technical failures are high lighted in a different colour...

Page 27: ...items are available from your local distributor or from ARGUS Medical AG 5 3 Software update procedure 5 3 1 General Please carefully check the software present installed on the pump If you have a fi...

Page 28: ...with firmware V4 xx With the Update center it is also possible to upgrade pump firmware older than V4 xx Important remark The actual calibration and configuration will be stored temporary on the PC p...

Page 29: ...by pressing the Download button 11 Important Perform a standard safety check see chap 10 the calibration values maybe lost during the upgrade procedure 5 4 Safety aspects Be aware of the following poi...

Page 30: ...every 24 month or after 10 000 hrs of operation The check has to be done in accordance to the chap ter 10 No special maintenance of the ARGUS 707 V infusion pump is necessary There are no wear and tea...

Page 31: ...r manual CAUTION The pressure calibration procedure always includes all of the follow ing steps including a control measurement Key functions during pressure calibration By the MODE key you can switch...

Page 32: ...P Display shows Set 1 2 Press START STOP Display shows CAL door 1 Press key 1 to select the set num ber you want to calibrate 1 to 4 6 3 2 Enter into the pressure calibration mode 6 3 3 Set selection...

Page 33: ...the downstream sensor 6 3 6 Calibrating the upstream sensor part 1 Remark By pressing key 1 the pressure signal in mV of the upstream sensor will be displayed after approximately 15 seconds the value...

Page 34: ...pressure build up When the pressure 1 2 bar reaches press START STOP 5 Release the pressure and press MODE Display shows CAL 1 0b 6 3 7 Calibrating the downstream sensor Remark The pump must reach a p...

Page 35: ...tomatically Is the pres sure of 0 2 bar reached press START STOP 7 Press MODE Display shows CAL 1 2b 5 Simulate an upstream occlusion by the 3 way stop cock 8 The pump continues on with the pres sure...

Page 36: ...anual Simulate an upstream occlusion by installing a clamping shears see chapter 6 3 6 or by kinking the tube To avoid the drop alarm release simulate falling drops on the drop detector The pump must...

Page 37: ...V set in the normal mode of the pump in menu Set x 1 Insert a new IV set only recommended IV sets may be used see appendix in the user manual in the pump and perform a warm up infusion of 40 ml at an...

Page 38: ...ion delivery time is 432 seconds It should now deliver a volume of 30 ml at an infusion rate of 250 ml h 7 After the delivery time has elapsed the pump stops and shows bAL 30 00 in the displays Enter...

Page 39: ...ressure sensor test failed always 4 7V or 0 2V delayed 5s F_39 Downstream pressure sensor test failed dynamical test failed F_40 Upstream pressure sensor test failed dynamical test failed F_44 Address...

Page 40: ...s and spare parts which have been approved by ARGUS Medical AG for safe technical use CAUTION If a door a housing a pressure sensor or a main board is replaced a full calibration pressure sensors and...

Page 41: ...s the centre Repeat point 1 and 2 for the other side of the door and then remove the door 1 Unscrew these screws Remove the pump unit 1 Use the screw with two washers at these places torque 1 0 Nm 2 U...

Page 42: ...rt touches nowhere otherwise noises may be generated 3 Reinstall the door Make sure that a washer has been installed on each side 5 Reconnect the back side with the front side battery power docking in...

Page 43: ...listed below What Where Torque in Nm Case Backplane into front 1 10 Pump unit Into front 1 00 Combination clamp Into clamp holder 1 50 Pressure sensor Into front 0 60 Stop flow Into front 0 60 Air de...

Page 44: ...parts 10 119 Battery NiMH 12V 1500mAh 10 087 Combination clamp 10 089 External drop detector 10 093 Interface cable docking pumps 10 120 Display board A707 10 121 Pressure sensor 10 122 Air detector...

Page 45: ...7 30VAC complete 10 136 Casing back plane A707 10 129 Casing forepart without door complete 10 137 Casing forepart 12 011 Interface cable 8pol 2m 12 012 Interface adapter 12 042 Cable staff alert 2m 1...

Page 46: ...l AG 28 03 06 PJ 45 48 12 035 Pressure gauge with stopcock 10 135 Edge board complete A707 11 005 Bottle holder 45 cm 11 043 Bottle holder 60 cm 12 044 Plug 11 219 11 239 11 247 Short instructions SW...

Page 47: ...e call EEPROM RTC Latch AC DC DC DC 100 10 0 1 Mode Start Stop DC DC 24V Motor Buzzer 5V Logic Controller Display Driver Motion Control Door Control Pressure Sensor in Pressure Sensor out 1 Motor Driv...

Page 48: ...ove any IV set Indicator air bubble lights up 9 Install a water filled IV set close the door Indicator air bubble turns off The red indicator drop lights up delayed 10 Set rate to 333 3 ml h press STA...

Page 49: ...4 ARGUS 100 P A400 ARGUS 100 M A404 ARGUS 600 S A200 ARGUS 707 V A300 ARGUS 708 V Serial Number Accessory Serial Number Production code Detailed failure or problem description Expected work repair to...

Reviews: