MAINTENANCE
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
29 / 48
6. MAINTENANCE
6.1. General
CAUTION!
Only authorized persons who have been trained by ARGUS Medical AG or by the
local distributor are allowed to service the ARGUS 707 V infusion pump. In case of
repair request, send the unit with the filled out “repair order form”
(see chapter 11)
to the local distributor. Further information is available from:
ARGUS Medical AG
CH-3627 Heimberg / Switzerland
E-mail:
CAUTION!
The safety standard check (SSC) has to be performed at least every 24 month or
after 10'000 hrs of operation. The check has to be done in accordance to the
chap-
ter 10
.
No special maintenance of the ARGUS 707 V infusion pump is necessary. There
are no wear and tear parts.
6.2. Recalibration
6.2.1. General
Unless otherwise specified by the customer, the ARGUS 707 V has been calibrated
by the manufacturer with the CODAN L86 PVC (
REF
43.4304) infusion set. If a dif-
ferent infusion set is used
(see recommended list in the appendix of the user man-
ual)
, a recalibration is required.
CAUTION!
A new set calibration always requires pressure sensors (up-/ downstream) and a
volume calibration
(see chapter 6.3 & 6.5)
!
It is mandatory to execute first the calibration procedure of the pressure sensors
and afterwards the volume calibration.