![Argus 707 V Service Manual Download Page 7](http://html1.mh-extra.com/html/argus/707-v/707-v_service-manual_2969492007.webp)
PUMP CONFIGURATIONS
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
6 / 48
5. Enter the PIN Code. If no PIN code
was configured, no entry is required
2. Press key “START”
3. Press key “MODE”
4. Press key “START” again.
The display will show “Cod ----“
2.3. Configuration mode (without ARGUS service)
The configuration mode permit you to modify the pump keypad configuration
manually using the keypad. Please refer to
chapter 2.6.
where the meaning of the
adresses are explained. To have access to all configuration options, please use the
“ARGUS
service
” PC-tool!
2.3.1. Step 1
2.3.2. Step 2
You have now access to all addresses in the list of
chapter 2.6.
Select therefore any address in the left display
(see next page)
.
Cod ----
707 p
1.
To enter into the configuration mode, switch the pump on
while keeping the keys “MODE” and “START/STOP” pressed.
6. Press key “START”
to acknowledge the
entered PIN code
If the PIN was not
accepted, the display
will change back to
point 5.
rog