Arctiko LTF 225 Operating Instructions Manual Download Page 9

9

PARAMETERS

DISPLAY

DEFROST

REGULATION

Hy

Differential:

 (0,1 ÷ 25,5°C / 1÷255 °F) Intervention differential for set point. Compressor Cut IN is 

Set Point + differential (Hy). Compressor Cut OUT is when the temperature reaches the set point.

LS

Minimum set point:

 (-100°C ÷ SET; [-148F ÷ SET): Sets the minimum value for the set point.

US

Maximum set point:

 (SET÷150°C/ SET÷302°F). Set the maximum value for set point.

Ot

Thermostat probe calibration:

 (-12.0÷12.0°C; -120÷120°F) allows to adjust possible offset of the 

thermostat probe.

OdS Outputs activation delay at start up:

 (0÷255min) This function is enabled at the initial start up of 

the instrument and inhibits any output activation for the period of time set in the parameter.

AC

Anti-short cycle delay:

 (0÷50 min) minimum interval between the compressor stop and the 

following restart.

CCt Compressor ON time during continuous cycle:

 (0.0÷24.0h; res. 10min) Allows to set the length of 

the continuous cycle: compressor stays on without interruption for the Cct time. Can be used, for 
instance, when the room is filled with new products.

CCS Set point for continuous cycle:

 (-100÷150°C; -148°F ÷ 302°F) it sets the set point used during the 

continuous cycle.

Con Compressor ON time with faulty probe:

 (0÷255 min) time during which the compressor is active in 

case of faulty thermostat probe. With COn=0 compressor is always OFF.

COF Compressor OFF time with faulty probe:

 (0÷255 min) time during which the compressor is OFF in 

case of faulty thermostat probe. With COF=0 compressor is always active.

CH

Type of action:

 CL = cooling; Ht = heating.

CF

Temperature measurement unit

°C

=Celsius; 

°F

=Fahrenheit. WARNING: When the measurement 

unit is changed the SET point and the values of the parameters Hy, LS, US, Ot, ALU and ALL have to 
be checked and modified if necessary).

rES Resolution (for °C):

 (in = 1°C; dE = 0.1 °C) allows decimal point display. 

dLy Display delay:

 (0 ÷20.0m; risul. 10s) when the temperature increases, the display is updated of 1 

°C/1°F after this time.

IdF

Interval between defrost cycles:

 (0÷120h) Determines the time interval between the beginning of 

two defrost cycles.

MdF (Maximum) length for defrost:

 (0÷255min) When

 P2P = n

, (not evaporator probe: timed defrost) it 

sets the defrost duration, when 

P2P = y

 (defrost end based on temperature) it sets the maximum 

length for defrost.

dFd Temperature displayed during defrost:

 (rt = real temperature; it = temperature at defrost start; 

SEt

 

= set point; dEF = 

“dEF”

 label)

dAd MAX display delay after defrost:

 (0÷255min). Sets the maximum time between the end of defrost 

and the restarting of the real room temperature display.

Summary of Contents for LTF 225

Page 1: ...OPERATING INSTRUCTIONS LTF 85 225 325 425 535...

Page 2: ...ioning Start Procedure Caution Control Panel Main Functions Parameters Digital Input Connections Default Setting Values Cleaning And Maintenance of The Freezer Relocating or Moving The Freezer Disposi...

Page 3: ...personnel in the use of the freezer the temperature is 60 C There is a risk of frost bite if the contents or the sides of the inner container are touched In the event of breakdown or irregularities co...

Page 4: ...locked in the freezer whilst playing Electrical Connection The machine meets current EU directives Low voltage 2006 95 E E C Electromagnetic compatibility 2004 108 E E C The machine should be given e...

Page 5: ...he installation should be fused with a 16 A fuse Before Commissioning Clean the freezer both inside and out using a damp wrung out cloth Dry with a dry cloth The smell from the plastic parts in the fr...

Page 6: ...emperature display USE OF LEDS Each LED function is described in the following table 1 Press and release the key 2 The Lo message will be displayed followed by the minimum temperature recorded 3 By pr...

Page 7: ...DEFROST Push the DEF key for more than 2 seconds and a manual defrost will start HOW TO CHANGE A PARAMETER VALUE To change the parameter s value operate as follows 1 Enter the Programming mode by pre...

Page 8: ...sage will be displayed and the keyboard will be locked At this point it will be possible only to see the set point or the MAX o Min temperature stored 3 If a key is pressed more than 3s the POF messag...

Page 9: ...n Compressor ON time with faulty probe 0 255 min time during which the compressor is active in case of faulty thermostat probe With COn 0 compressor is always OFF COF Compressor OFF time with faulty p...

Page 10: ...r failure y the alarm relay is activated if a temperature alarm happens during a power failure n the alarm relay is never activated during a power failure ALF Alarm relay activation for all the alarms...

Page 11: ...1F parameter DOOR SWITCH INPUT i1F dor It signals the door status and the corresponding relay output status through the odc parameter no Fan normal any change CPr F_C Compressor OFF Since the door is...

Page 12: ...unction is enabled until the digital input is activated DIGITAL INPUTS POLARITY The digital input polarity depends on the i1P parameter i1P CL the input is activated by closing the contact i1P OP the...

Page 13: ...13 CONNECTIONS XR30CX 20A COMPRESSOR 12Vac dc supply connect to the terminals 7 and 8 24Vac dc supply connect to the terminals 7 and 8 120Vac supply connect to the terminals 7 and 8...

Page 14: ...n rE related to set Ab absolute rE Pr2 ALU Maximum temperature alarm Set 110 0 C Set 230 F 15 Pr1 ALL Minimum temperature alarm 50 0 C Set 58 F Set 50 0 Pr1 AFH Differential for temperature alarm reco...

Page 15: ...reful not to get damp water on the electrical parts or water into the vent holes as water and damp may damage the electrical parts and cause the freezer to short circuit 4 The ice formed on the inside...

Page 16: ...rting it If the freezer and its electrical parts are reconstructed the guarantee and safety regulations shall cease to apply If the freezer does not work call an authorised servicing engineer Contact...

Page 17: ...17 Wiring Diagram LTF 85 325 535...

Page 18: ...18 Wiring Diagram LTF 225...

Page 19: ...19 Wiring Diagram LTF 425...

Page 20: ...50 80 017 04 GB WWW ARCTIKO COM LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 INFO ARCTIKO COM WWW ARCTIKO COM...

Reviews: