Arctiko LTF 225 Operating Instructions Manual Download Page 8

8

3. Select the required parameter.
4. Press the 

“SET

” key to display its value.

5. Use        or        to change its value.
6. Press 

“SET”

 to store the new value and move to the following parameter. 

To exit:

 Press 

SET +

        or 

wait 15s without pressing a key.

NOTE1:

if none parameter is present in Pr1, after 3s the “noP” message is displayed. Keep the keys 
pushed till the Pr2 message is displayed.

NOTE2:

the set value is stored even when the procedure is exited by waiting the time-out to expire.

6.6.2 HOW TO MOVE A PARAMETER FROM THE HIDDEN MENU TO THE FIRST LEVEL AND 

VICEVERSA.

Each parameter present in the HIDDEN MENU can be removed or put into “THE FIRST LEVEL” (user level) 

by pressing 

“SET”

 + 

In HIDDEN MENU when a parameter is present in First Level the decimal point is on.

HOW TO LOCK THE KEYBOARD

1. Keep pressed for more than 3 s the        and         keys.
2. The 

“POF”

 message will be displayed and the keyboard will be locked. At this point it will be possible only 

to see the set point or the MAX o Min temperature stored

3. If a key is pressed more than 3s the “POF” message will be displayed.

TO UNLOCK THE KEYBOARD

Keep pressed together for more than 3s the        and         keys, till the “Pon” message will be displayed.

THE CONTINUOUS CYCLE

When defrost is not in progress, it can be activated by holding the         key pressed  .

for about 3 seconds.

The compressor operates to maintain the 

“ccS”

 set point for the time set through the 

“CCt”

 parameter. The 

cycle can be terminated before the end of the set time using the same activation key “         ” for 3 seconds.

THE ON/OFF FUNCTION

With 

“onF = oFF”

, pushing the 

ON/OFF

 key, the instrument is switched off. The 

“OFF”

 message is displayed. 

In this configuration, the regulation is disabled. To switch the instrument on, push again the ON/OFF key.

WARNING: Loads connected to the normally closed contacts of the relays are always supplied and 

under voltage, even if the instrument is in stand by mode.

Summary of Contents for LTF 225

Page 1: ...OPERATING INSTRUCTIONS LTF 85 225 325 425 535...

Page 2: ...ioning Start Procedure Caution Control Panel Main Functions Parameters Digital Input Connections Default Setting Values Cleaning And Maintenance of The Freezer Relocating or Moving The Freezer Disposi...

Page 3: ...personnel in the use of the freezer the temperature is 60 C There is a risk of frost bite if the contents or the sides of the inner container are touched In the event of breakdown or irregularities co...

Page 4: ...locked in the freezer whilst playing Electrical Connection The machine meets current EU directives Low voltage 2006 95 E E C Electromagnetic compatibility 2004 108 E E C The machine should be given e...

Page 5: ...he installation should be fused with a 16 A fuse Before Commissioning Clean the freezer both inside and out using a damp wrung out cloth Dry with a dry cloth The smell from the plastic parts in the fr...

Page 6: ...emperature display USE OF LEDS Each LED function is described in the following table 1 Press and release the key 2 The Lo message will be displayed followed by the minimum temperature recorded 3 By pr...

Page 7: ...DEFROST Push the DEF key for more than 2 seconds and a manual defrost will start HOW TO CHANGE A PARAMETER VALUE To change the parameter s value operate as follows 1 Enter the Programming mode by pre...

Page 8: ...sage will be displayed and the keyboard will be locked At this point it will be possible only to see the set point or the MAX o Min temperature stored 3 If a key is pressed more than 3s the POF messag...

Page 9: ...n Compressor ON time with faulty probe 0 255 min time during which the compressor is active in case of faulty thermostat probe With COn 0 compressor is always OFF COF Compressor OFF time with faulty p...

Page 10: ...r failure y the alarm relay is activated if a temperature alarm happens during a power failure n the alarm relay is never activated during a power failure ALF Alarm relay activation for all the alarms...

Page 11: ...1F parameter DOOR SWITCH INPUT i1F dor It signals the door status and the corresponding relay output status through the odc parameter no Fan normal any change CPr F_C Compressor OFF Since the door is...

Page 12: ...unction is enabled until the digital input is activated DIGITAL INPUTS POLARITY The digital input polarity depends on the i1P parameter i1P CL the input is activated by closing the contact i1P OP the...

Page 13: ...13 CONNECTIONS XR30CX 20A COMPRESSOR 12Vac dc supply connect to the terminals 7 and 8 24Vac dc supply connect to the terminals 7 and 8 120Vac supply connect to the terminals 7 and 8...

Page 14: ...n rE related to set Ab absolute rE Pr2 ALU Maximum temperature alarm Set 110 0 C Set 230 F 15 Pr1 ALL Minimum temperature alarm 50 0 C Set 58 F Set 50 0 Pr1 AFH Differential for temperature alarm reco...

Page 15: ...reful not to get damp water on the electrical parts or water into the vent holes as water and damp may damage the electrical parts and cause the freezer to short circuit 4 The ice formed on the inside...

Page 16: ...rting it If the freezer and its electrical parts are reconstructed the guarantee and safety regulations shall cease to apply If the freezer does not work call an authorised servicing engineer Contact...

Page 17: ...17 Wiring Diagram LTF 85 325 535...

Page 18: ...18 Wiring Diagram LTF 225...

Page 19: ...19 Wiring Diagram LTF 425...

Page 20: ...50 80 017 04 GB WWW ARCTIKO COM LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 INFO ARCTIKO COM WWW ARCTIKO COM...

Reviews: