Arctiko LTF 225 Operating Instructions Manual Download Page 14

14

DEFAULT SETTING VALUES 

Label 

Name 

Range 

Set point  Pr 1/2 

Set 

Se

t

 point 

LS÷US 

-62 

--- 

Hy 

Differential 

0,1÷25,5°C/1÷255°F 

Pr1 

LS 

Minimum set point 

-50°C÷SET/-58°F÷SET 

-63 

Pr2 

US 

Maximum set point 

SET÷110°C/SET÷230°F 

-30 

Pr2 

Ot 

Thermostat probe calibration 

-12÷12°C/-120÷120°F 

Pr1 

OdS 

Outputs delay at start up 

0÷255min 

Pr2 

AC 

Anti-short cycle delay 

0÷50min 

Pr1 

CCt 

Continuous cycle duration 

0.0÷24.0h 

Pr2 

CCS 

Set point for continuous cycle 

(-55.0÷150.0°C) (-67÷302°F) 

-62 

Pr2 

COn 

Compressor ON time with faulty probe 

0÷255min 

60 

Pr2 

COF 

Compressor OFF time with faulty probe 

÷255min 

Pr2 

CH 

Kind of action 

CL=cooling; Ht=heating 

Cl 

Pr1 

CF 

Temperature measurement unit 

°C÷°F 

°C 

Pr2 

rES 

Resolution 

In=integer; dE=dec.point 

in 

Pr1 

dLy 

Display temperature delay 

0÷20.0min (10sec.) 

Pr2 

IdF 

Interval between defrost cycles 

1÷120ore 

Pr1 

MdF 

(Maximum) length for defrost 

0÷255min 

Pr1 

dFd 

Displaying duing defrost 

rt, it, Set, DEF 

rt 

Pr2 

dAd 

MAX display delay after defrost 

0÷255min 

Pr2 

ALc 

Temperature alarms configuration 

rE=related to set;  Ab=absolute 

rE 

Pr2 

ALU 

Maximum temperature alarm 

Set÷110.0°C;  Set÷230°F 

15 

Pr1 

ALL 

Minimum temperature alarm 

-50.0°C÷Set/-58°F÷Set 

-50.0 

Pr1 

AFH 

Differential for temperature alarm 
recovery 

(0.1°C÷25.5°C)(1°F÷45°F) 

Pr2 

ALd 

Temperature alarm delay 

0÷255min 

Pr2 

dAo 

Delay of temperature alarm at start up 

0÷23h e 50’ 

Pr2 

tbA 

Alarm relay disabling 

n=no; y=yes 

Pr2 

Aro 

Alarm relay activation with power failure 

n(0) – y(1) 

Pr2 

ALF 

Alarm relay activation for all the alarms 

n(0) – y(1) 

Pr2 

bon 

Time of buzzer restart after muting, in 
case of alarm duration 

0 ÷ 30 (min) 

30 

Pr2 

AoP 

Alarm relay polarity (0A1=ALr) 

oP; cL 

cL 

Pr2 

i1P 

Digital input polarity 

oP=opening; cL=closing 

cL 

Pr1 

i1F 

Digital input configuration 

EAL, bAL,PAL, dor; dEF; Htr, 

AUS 

AUS 

Pr1 

did 

Digital input alarm delay 

0÷255min 

15 

Pr1 

nPS 

Number of activation of pressure switch 

0÷15 

15 

Pr2 

odc 

Compress status when open door 

No; Fan;CPr;F_C 

no 

Pr2 

rrd 

Regulation restart with door open alarm 

n - y 

Pr2 

HES 

Differential for Energy Saving 

(-30°C÷30°C) (-54°F÷54°F) 

Pr2 

Adr 

Serial address 

0÷247 

Pr2 

PbC 

Kind of probe 

Ptc; ntc 

Pt1 

Pr2 

onF 

On/off key enabling 

Nu, oFF; ES 

nu 

Pr2 

rSE 

Real set point value 

Actual set 

--- 

Pr2 

rEL 

Software release 

--- 

--- 

Pr2 

0

OFF

Summary of Contents for LTF 225

Page 1: ...OPERATING INSTRUCTIONS LTF 85 225 325 425 535...

Page 2: ...ioning Start Procedure Caution Control Panel Main Functions Parameters Digital Input Connections Default Setting Values Cleaning And Maintenance of The Freezer Relocating or Moving The Freezer Disposi...

Page 3: ...personnel in the use of the freezer the temperature is 60 C There is a risk of frost bite if the contents or the sides of the inner container are touched In the event of breakdown or irregularities co...

Page 4: ...locked in the freezer whilst playing Electrical Connection The machine meets current EU directives Low voltage 2006 95 E E C Electromagnetic compatibility 2004 108 E E C The machine should be given e...

Page 5: ...he installation should be fused with a 16 A fuse Before Commissioning Clean the freezer both inside and out using a damp wrung out cloth Dry with a dry cloth The smell from the plastic parts in the fr...

Page 6: ...emperature display USE OF LEDS Each LED function is described in the following table 1 Press and release the key 2 The Lo message will be displayed followed by the minimum temperature recorded 3 By pr...

Page 7: ...DEFROST Push the DEF key for more than 2 seconds and a manual defrost will start HOW TO CHANGE A PARAMETER VALUE To change the parameter s value operate as follows 1 Enter the Programming mode by pre...

Page 8: ...sage will be displayed and the keyboard will be locked At this point it will be possible only to see the set point or the MAX o Min temperature stored 3 If a key is pressed more than 3s the POF messag...

Page 9: ...n Compressor ON time with faulty probe 0 255 min time during which the compressor is active in case of faulty thermostat probe With COn 0 compressor is always OFF COF Compressor OFF time with faulty p...

Page 10: ...r failure y the alarm relay is activated if a temperature alarm happens during a power failure n the alarm relay is never activated during a power failure ALF Alarm relay activation for all the alarms...

Page 11: ...1F parameter DOOR SWITCH INPUT i1F dor It signals the door status and the corresponding relay output status through the odc parameter no Fan normal any change CPr F_C Compressor OFF Since the door is...

Page 12: ...unction is enabled until the digital input is activated DIGITAL INPUTS POLARITY The digital input polarity depends on the i1P parameter i1P CL the input is activated by closing the contact i1P OP the...

Page 13: ...13 CONNECTIONS XR30CX 20A COMPRESSOR 12Vac dc supply connect to the terminals 7 and 8 24Vac dc supply connect to the terminals 7 and 8 120Vac supply connect to the terminals 7 and 8...

Page 14: ...n rE related to set Ab absolute rE Pr2 ALU Maximum temperature alarm Set 110 0 C Set 230 F 15 Pr1 ALL Minimum temperature alarm 50 0 C Set 58 F Set 50 0 Pr1 AFH Differential for temperature alarm reco...

Page 15: ...reful not to get damp water on the electrical parts or water into the vent holes as water and damp may damage the electrical parts and cause the freezer to short circuit 4 The ice formed on the inside...

Page 16: ...rting it If the freezer and its electrical parts are reconstructed the guarantee and safety regulations shall cease to apply If the freezer does not work call an authorised servicing engineer Contact...

Page 17: ...17 Wiring Diagram LTF 85 325 535...

Page 18: ...18 Wiring Diagram LTF 225...

Page 19: ...19 Wiring Diagram LTF 425...

Page 20: ...50 80 017 04 GB WWW ARCTIKO COM LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 INFO ARCTIKO COM WWW ARCTIKO COM...

Reviews: