Arctiko LTF 225 Operating Instructions Manual Download Page 11

11

nPS Pressure switch number

: (0 ÷15) Number of activation of the pressure switch, during the “did” interval, 

before signalling the alarm event (I2F= PAL). If the nPS activation in the did time is reached, switch off 
and on the instrument to
restart normal regulation.

odc Compressor status with door open:

 no, Fan = normal; CPr; 

F_C =

 Compressor OFF.

rrd

Outputs restart after doA alarm:

 

no =

 outputs not affected by the doA alarm;

 yES =

 outputs restart 

with the doA alarm;

HES Temperature increase during the Energy Saving cycle:

 (-30,0°C÷30,0°C/-22÷86°F) it sets the 

increasing value of the set point during the Energy Saving cycle.

Adr Serial address

 (1÷244): Identifies the instrument address when connected to a ModBUS compatible 

monitoring system.

PbC Type of probe:

 it allows to set the kind of probe used by the instrument: PtC = PTC probe, Pt1 = 

Pt1000 probe.

onF on/off key enabling:

 nu = disabled; oFF = enabled; ES = not set it.

rSE Real set point:

 (readable only), it shows the set point used during the energy saving cycle or during the 

continuous cycle.

rEL Software release

 for internal use.

Ptb Parameter table code:

 readable only.

DIGITAL INPUT

 The free voltage digital input is programmable in different configurations by the “i1F” parameter.

DOOR SWITCH INPUT (i1F = dor)

It signals the door status and the corresponding relay output status through the “odc” parameter: 

no, Fan =

 

normal (any change); 

CPr, F_C =

 Compressor OFF.

Since the door is opened, after the delay time set through parameter “did”, the door alarm is enabled, the 
display shows the message 

“dA” and the regulation restarts is rtr = yES

. The alarm stops as soon as 

the external digital input is disabled again. With the door open, the high and low temperature alarms are 
disabled.

GENERIC ALARM (i1F = EAL)

As soon as the digital input is activated the unit will wait for 

“did”

 time delay before signalling the

 “EAL”

 

alarm message. The outputs status don’t change. The alarm stops just after the digital input is de-activated.

SERIOUS ALARM MODE (i1F = bAL)

When the digital input is activated, the unit will wait for 

“did”

 delay before signalling the 

“CA”

 alarm 

message. The relay outputs are switched OFF. The alarm will stop as soon as the digital input is de-
activated.

OTHER

PRESSURE SWITCH (i1F = PAL)

If during the interval time set by 

“did”

 parameter, the pressure switch has reached the number of activation 

of the 

“nPS”

 parameter, the 

“CA”

 pressure alarm message will be displayed. The compressor and the 

regulation are stopped. When the digital input is ON the compressor is always OFF.

If the nPS activation in the did time is reached, switch off and on the instrument to restart normal 

regulation.

Summary of Contents for LTF 225

Page 1: ...OPERATING INSTRUCTIONS LTF 85 225 325 425 535...

Page 2: ...ioning Start Procedure Caution Control Panel Main Functions Parameters Digital Input Connections Default Setting Values Cleaning And Maintenance of The Freezer Relocating or Moving The Freezer Disposi...

Page 3: ...personnel in the use of the freezer the temperature is 60 C There is a risk of frost bite if the contents or the sides of the inner container are touched In the event of breakdown or irregularities co...

Page 4: ...locked in the freezer whilst playing Electrical Connection The machine meets current EU directives Low voltage 2006 95 E E C Electromagnetic compatibility 2004 108 E E C The machine should be given e...

Page 5: ...he installation should be fused with a 16 A fuse Before Commissioning Clean the freezer both inside and out using a damp wrung out cloth Dry with a dry cloth The smell from the plastic parts in the fr...

Page 6: ...emperature display USE OF LEDS Each LED function is described in the following table 1 Press and release the key 2 The Lo message will be displayed followed by the minimum temperature recorded 3 By pr...

Page 7: ...DEFROST Push the DEF key for more than 2 seconds and a manual defrost will start HOW TO CHANGE A PARAMETER VALUE To change the parameter s value operate as follows 1 Enter the Programming mode by pre...

Page 8: ...sage will be displayed and the keyboard will be locked At this point it will be possible only to see the set point or the MAX o Min temperature stored 3 If a key is pressed more than 3s the POF messag...

Page 9: ...n Compressor ON time with faulty probe 0 255 min time during which the compressor is active in case of faulty thermostat probe With COn 0 compressor is always OFF COF Compressor OFF time with faulty p...

Page 10: ...r failure y the alarm relay is activated if a temperature alarm happens during a power failure n the alarm relay is never activated during a power failure ALF Alarm relay activation for all the alarms...

Page 11: ...1F parameter DOOR SWITCH INPUT i1F dor It signals the door status and the corresponding relay output status through the odc parameter no Fan normal any change CPr F_C Compressor OFF Since the door is...

Page 12: ...unction is enabled until the digital input is activated DIGITAL INPUTS POLARITY The digital input polarity depends on the i1P parameter i1P CL the input is activated by closing the contact i1P OP the...

Page 13: ...13 CONNECTIONS XR30CX 20A COMPRESSOR 12Vac dc supply connect to the terminals 7 and 8 24Vac dc supply connect to the terminals 7 and 8 120Vac supply connect to the terminals 7 and 8...

Page 14: ...n rE related to set Ab absolute rE Pr2 ALU Maximum temperature alarm Set 110 0 C Set 230 F 15 Pr1 ALL Minimum temperature alarm 50 0 C Set 58 F Set 50 0 Pr1 AFH Differential for temperature alarm reco...

Page 15: ...reful not to get damp water on the electrical parts or water into the vent holes as water and damp may damage the electrical parts and cause the freezer to short circuit 4 The ice formed on the inside...

Page 16: ...rting it If the freezer and its electrical parts are reconstructed the guarantee and safety regulations shall cease to apply If the freezer does not work call an authorised servicing engineer Contact...

Page 17: ...17 Wiring Diagram LTF 85 325 535...

Page 18: ...18 Wiring Diagram LTF 225...

Page 19: ...19 Wiring Diagram LTF 425...

Page 20: ...50 80 017 04 GB WWW ARCTIKO COM LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 INFO ARCTIKO COM WWW ARCTIKO COM...

Reviews: