39
In the
pH calibration
menu item, start the mode by
pressing the
OK
button.
Dip the pH probe into the pH 7 buffer solution and
start by pressing the
OK
button.
Press
OK
again to trigger the countdown
After 20 seconds, the control unit shows the value of
the used buffer solution and the slope in mV. If the
calibration is valid, continue by pressing
OK
Then take the probe from the first buffer solution and,
if possible, rinse with water and dry with a dry, lint-free
paper towel.
Caution:
Do not rub the probe to dry it since this causes electrostatic charge and can
result in incorrectly measured values.
Now dip the probe into the second pH 4 buffer solution.
Press
OK
again to trigger the countdown.
After another 20 seconds, the control unit shows the
values
of
the
second
buffer
solution
pH
4.
If the calibration was valid, continue by pressing
OK
The display appears with the specification of the slope
in mV
Go back using
Mode
or accept using
OK
.
The control unit has now been successfully calibrated to
the pH probe and the data has been saved.
If the calibration process was not successful, "Error" is
displayed. Cancel using
Mode
or repeat using
OK
.
There are a number of potential causes of
calibration errors
:
•
You have used the same buffer solution twice during calibration. Correct
calibration can only be carried out with two different technical buffer solutions.
pH 7 first and then pH 4.
•
The measuring cable was connected incorrectly. The pH probe must be connected
to the left socket.
Calibration of probe
Calibration of probe
pH7
Calibration of probe
(20)
Calibration of probe
pH4
Calibration of probe
(20)
Calibration of probe
successful
Error value!
OK
OK
OK
Summary of Contents for SL3 Deluxe Wi-Fi
Page 146: ...146 RU Aqua Consulting Swim tec DOS SL 3 Deluxe Wi Fi IDS pH 2 4 I II...
Page 148: ...148 1 1 2 3 4 pH 5 6 pH 7 8 1 3 5 6 4 7 8 2...
Page 150: ...150 Flow Level Limit OK OK 5 C Min Alarm Max Alarm pH 4 4 4 5 pH...
Page 152: ...152 Mode 3 3 1 OK STOP pH OK 3 2 1 2 3 MODE OK 1 2 3...
Page 153: ...153 pH pH ein aus OK 2 100 3 3 3 3 1 3 3 2 3 3 3 3 3 4 3 4 3 5 IDS IDS...
Page 156: ...156 OK Mode OK pH Mode OK pH 7 pH 4 pH pH...
Page 157: ...157 4 2 pH 7 2 pH pH 4 3 pH 1 0 pH 1 00 pH 1 00 pH pH pH SN6 pH pH...
Page 158: ...158 1 00 1 00 1 00 1 0 5 50 4 4 Min Alarm 6 80 4 5 Max Alarm 7 60 5 pH 28 C 5 1...
Page 162: ...162 7 5 7 6 MAC MAC 7 7 VPN VPN Apple iOS iPhone iPad iPod touch Android DOS Deluxe Wifi VPN...
Page 165: ...165 1 1 1 65 1 2...
Page 166: ...166 2 VDE 0105 230 VDE DIN VDE 0100 702 VDE DIN VDE 0100 702 3 I N 30...
Page 167: ...167 DIN EN 809 3 5 40 C...
Page 168: ...168 4 4 1 20 pH 8 5 6 2 3 10 7 0 5 9 4 1 11 1 1 5 9 2 6 10 3 7 pH 11 4 8 i 9...
Page 169: ...169 4 2 pH 5 5 1...
Page 170: ...170 5 1 1 5 1 2 pH 5 1 3 pH 5 2 5 2 1 3...
Page 171: ...171 5 2 2 pH 5 2 3 0 1 3 1 5 5 2 4...
Page 173: ...173 Pharmed 5 4 5 4 1 5 4 2 3 PG13...
Page 176: ...176...