59
14.
Step Cycle (PROGRAM Mode only)
- gives the operator the flexibility to
program the number of looping cycles for a particular step. For example if the
operator would like to have a step repeat five times the step cycle would be
programmed to five. Step Cycle is only selectable when the source is
configured for AC output.
15.
A Hi-Lmt (PROGRAM Mode)
- gives the operator the flexibility to program
a maximum current threshold or ceiling level. When this level is reached a
failure will occur.
16.
A Lo-Lmt (PROGRAM Mode only)
- gives the operator the flexibility to
program a minimum current threshold or floor level. If a minimum current
level is not reached a failure will occur. This insures a load is attached to the
power source and there is a minimum current present.
17.
P Hi-Lmt (PROGRAM Mode only)
- gives the operator the flexibility to
program a maximum wattage threshold or ceiling level. When this level is
reached a failure will occur.
18.
P Lo-Lmt (PROGRAM Mode only)
- gives the operator the flexibility to
program a minimum wattage threshold or floor level. If a minimum wattage
level is not reached a failure will occur. This insures a load is attached to the
power source and there is a minimum wattage present.
19.
Ap Hi-Lmt (PROGRAM Mode only)
- gives the operator the flexibility to
program a maximum peak current threshold or ceiling level. When this level
is reached a failure will occur. Ap Hi-Lmt is only selectable when the source
is configured for AC output.
20.
Ap Lo-Lmt (PROGRAM Mode only)
- gives the operator the flexibility to
program a minimum peak current threshold or floor level. If a minimum peak
current level is not reached a failure will occur. This ensures a load is
attached to the power source and there is a minimum peak current present.
Ap Lo-Lmt is only selectable when the source is configured for AC output.
21.
CF Hi-Lmt (PROGRAM Mode only)
- gives the operator the flexibility to
program a maximum crest factor threshold or floor level. When this level is
reach a failure will occur. CF Hi-Lmt is only selectable when the source is
configured for AC output.
22.
CF Lo-Lmt (PROGRAM Mode only)
- gives the operator the flexibility to
program a minimum crest factor threshold or floor level. If a minimum crest
factor level is not reached a failure will occur. CF Lo-Lmt is only selectable
when the source is configured for AC output.