2.2.2 Selecting the Appropriate Wire Gauge
Below is a table which provides recommended appropriate wire gauges for the
300XAC.
310XAC
Input
Output
Terminal
Line
Neutral
Ground
Line
Neutral
Ground
Nsense
Lsense
Gauge
12 AWG
12 AWG
12
AWG 14
AWG 14
AWG 14
AWG 18
AWG 18
AWG
320XAC
Input
Output
Terminal
Line
Neutral
Ground
Line
Neutral
Ground
Nsense
Lsense
Gauge
12 AWG
12 AWG
12
AWG 14
AWG 14
AWG 14
AWG 18
AWG 18
AWG
340XAC
Input
Output
Terminal
Line
Neutral
Ground
Line
Neutral
Ground
Nsense
Lsense
Gauge
10 AWG
10 AWG
10
AWG 10
AWG 10
AWG 10
AWG 18
AWG 18
AWG
360XAC
Input
Output
Terminal
Line(A,B,C) Neutral
(A,B,C) Ground
Line
Neutral
Ground
Nsense
Lsense
2W
10 AWG
10 AWG
10 AWG
3W
12 AWG
N/A
12 AWG
4W
14 AWG
14 AWG
14 AWG
8 AWG
8 AWG
8
AWG
18
AWG 18
AWG
2.2.3 Power Cable
Before connecting power to this instrument, the protective
ground (earth) terminals of this instrument must be connected to the protective
conductor of the line (mains) power cord. The main plug shall only be inserted in
a socket outlet (receptacle) provided with a protective ground (earth) contact.
This protective ground (earth) must not be defeated by the use of an extension
cord without a protective conductor (grounding).
360XAC Power Cable Considerations
The 360XAC can be powered with several different input power configurations.
Refer to the diagram below for connecting a single phase, 3 phase wye, and 3
phase delta power source to the instrument:
WARNING