2
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]
Performance
Operating Range / Transducer Frequency
Option 2 - 69 KHz (-20 - 50°C)
1.0 - 25.0 ft. / 12 in. - 300 in. / 305mm - 7620mm
Option 6 - 69 KHz (-30 - 60°C)
1.0 - 25.0 ft. / 12 in. - 300 in. / 305mm - 7620mm
Digital Output
Modbus, via RS-485
Beam Pattern
9° off axis
Sample Rate
1 - 20 Hz
Response Time
Programmable
Accuracy
Accuracy
Greater of ±0.25% of detected range or ± 3mm
Resolution
0.1
inch
(2.54
mm)
Environmental
Operating & Storage Temperature
Ambient -30 to 60°C (-22 to 140°F)
unless otherwise specified
Internal Temperature Compensation
Yes
Humidity 100%
non-condensing
IP rating
IP68, submerged 3 meters for 48 hours
Electrical
Supply Voltage (at sensor)
8-24 VDC
Opperating Current Draw (dependent on setup)
@ 8 VDC:
Min = 30 mA, Max = 75 mA rms
@ 24 VDC:
Min = 15 mA, Max = 30 mA rms
Waiting Status Current Draw
@
8
VDC: 17
mA
rms
@
24
VDC:
13
mA
rms
Connection
2 Shielded Twisted Pair and drain wire
Materials of Construction
Transducer Face
Housing
Option
2
Kynar
PVDF
Housing PBT/Polycarbonate
blend
Mounting
Housing Option 2
2” NPT (Housing body), 3/4” NPT (Lid)
IS Entity Parameters
See Hazard Drawing 9005002 at back of manual for IS Entity Parameter listing.
•
Specifications