background image

 

- 2 -

 

Inhaltsverzeichnis 

1.

 

Introduction

 ------------------------------------------------------------------------------------------------------- 

- 3 -

 

2.

 

What’s in the Kit

 ------------------------------------------------------------------------------------------------- 

- 4 -

 

3.

 

Battery Installation

 ---------------------------------------------------------------------------------------------- 

- 4 -

 

4.

 

Keypad Functions

 ------------------------------------------------------------------------------------------------ 

- 5 -

 

5.

 

Preparation before Use

 ---------------------------------------------------------------------------------------- 

- 5 -

 

6.

 

pH Calibration

 ---------------------------------------------------------------------------------------------------- 

- 5 -

 

7.

 

pH Measurement

 ------------------------------------------------------------------------------------------------ 

- 7 -

 

8.

 

ORP Measurement

 ---------------------------------------------------------------------------------------------- 

- 8 -

 

9.

 

Probe Cleaning

 --------------------------------------------------------------------------------------------------- 

- 8 -

 

10.

 

Probe Storage

 -------------------------------------------------------------------------------------------------- 

- 8 -

 

11.

 

Parameter Setting

 --------------------------------------------------------------------------------------------- 

- 9 -

 

12.

 

Technical Specifications

 ----------------------------------------------------------------------------------- 

- 10 -

 

13.

 

Icons and Functions

 ---------------------------------------------------------------------------------------- 

- 10 -

 

14.

 

Probe Replacement

 ---------------------------------------------------------------------------------------- 

- 11 -

 

15.

 

Troubleshooting Guide

 ------------------------------------------------------------------------------------ 

- 11 -

 

16.

 

Warranty

 ------------------------------------------------------------------------------------------------------ 

- 12 -

 

 

 

 

ATTENTION

 

 

Never

 

use the product when it’s freezing cold. Let it warm to room temperature before using.

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for AI3711

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...Calibration 5 7 pH Measurement 7 8 ORP Measurement 8 9 Probe Cleaning 8 10 Probe Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Tr...

Page 3: ...connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to sw...

Page 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Page 5: ...calibration bottles to about half volume of the bottle Long press to enter calibration mode screen turns green Insert the probe into pH 7 00 buffer solution make a quick stir and hold still When the r...

Page 6: ...H first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 10 times as long as they are not contaminated and the bottles are capped when not in use pH 10 0...

Page 7: ...ibration again 3 The electrode is aged used for over a year and has a much slower response A replacement is needed pH electrode zero electric potential out of range 60mV or 60mV 1 Check whether pH buf...

Page 8: ...d to be able to measure ORP 2 Power on the PC60 Z tester unscrew the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14...

Page 9: ...rt press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut down af...

Page 10: ...ature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV Accuracy 0 2 F S Temperature Range 0 to 50 C 32 122 F Accuracy 0 5 C Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6...

Page 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Page 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...