background image

 

- 3 -

 

1.  Introduction 

Thank you for choosing Apera Instruments PH60-Z Smart pH Tester. Please carefully read this manual 
before using the product.

 

 

 

Search “zentest smart” in Apple App Store or Google Play App Store to download the latest App

 

for your tester. Turn on the Bluetooth of your phone, go to ZenTest™ App, tap 

on the upper 

right corner, then select your tester to connect. 

  For video tutorials on how to get the most out of 

ZenTest™

, please go to 

www.aperainst.de

 

  This product is designed with a two-

way control on both the tester and ZenTest™ App. Please 

refer to the functions available on each platform in the following table. This manual shows you how 

to operate the tester without connecting to a smartphone.   

 

Table 1: Funtions on 60-Z Tester and ZenTest

® 

Mobile App

 

Functions

 

60-Z Tester

 

ZenTest Mobile App

 

Display

 

LCD display

 

1. 

Basic Mode: digital display 
+calibration info

 

Swipe to 
switch 
among 
various 
modes

 

2. 

Dial Mode: digital dial 
display

 

3. 

Graph Mode: digital graph 
display

 

4. 

Table Mode: digital real time 
measurement and history display

 

Calibration

 

Press buttons to 
operate

 

Operate on smartphone following graphic guides

 

Self-
Diagnosis

 

Er1 – Er6 icons

 

Detailed problem analysis and solutions

 

Parameter 
Setup

 

Press buttons to set up 
(except for P7 and P11)

 

All parameters can be set up in Settings.

 

Alarm

 

The screen turns red 
when alarm triggered; 
cannot be setup

 

Alarm display and alarm values can be preset for each 
parameter

 

Datalogger

 

N/A

 

Manual or Auto. Datalogger; notes can be added to 
saved data

 

Data 
Output

 

N/A

 

Share data via Email

 

Summary of Contents for AI3711

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...Calibration 5 7 pH Measurement 7 8 ORP Measurement 8 9 Probe Cleaning 8 10 Probe Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Tr...

Page 3: ...connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to sw...

Page 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Page 5: ...calibration bottles to about half volume of the bottle Long press to enter calibration mode screen turns green Insert the probe into pH 7 00 buffer solution make a quick stir and hold still When the r...

Page 6: ...H first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 10 times as long as they are not contaminated and the bottles are capped when not in use pH 10 0...

Page 7: ...ibration again 3 The electrode is aged used for over a year and has a much slower response A replacement is needed pH electrode zero electric potential out of range 60mV or 60mV 1 Check whether pH buf...

Page 8: ...d to be able to measure ORP 2 Power on the PC60 Z tester unscrew the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14...

Page 9: ...rt press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut down af...

Page 10: ...ature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV Accuracy 0 2 F S Temperature Range 0 to 50 C 32 122 F Accuracy 0 5 C Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6...

Page 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Page 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...

Reviews: