- 9 -

 

11. Parameter Setting 

 Table of Settings 

 

 Parameter Setting 

1) 

When the meter is turned off, long press 

  to enter parameter setting →short press 

  to 

switch  P01-P02…  →P8.  Short  Press 

 ,  parameter  flashes  →  short  press 

   to  adjust 

parameter → short press 

  to confirm →Short press 

  to exit parameter setting and go back 

to measurement mode.

 

2) 

Auto. Lock (P02)

 – Users can set the auto lock time from 5 to 20 seconds. For example, if 10 

seconds is set, when the measured value is stable for more than 10 seconds, the measured value 
will be automatically locked, and the HOLD icon will be displayed. Short press 

to release the 

lock. When the setting is "Off", the Auto. lock function is turned off, that is, the measured value can 
only be manually locked. Short press 

to lock or unlock the measured value. The HOLD icon 

will be displayed when reading is locked.

 

3) 

Auto.  Backlight  (P03)

 

  Users  can  set  the  automatic  backlight  time  for  1  to  8  minutes.  For 

example, if 3 minutes is set, the backlight will turn off automatically after 3 minutes; when the “Off” 
is  set,  the  auto.  backlight  function  will  be  turned  off,  and  short  press 

 to  manually  turn  the 

backlight on or off.

 

4) 

Auto. Power off (P04)

 

  The auto. power off time can be set to 10 to 20 minutes. For example, 

if 15 minutes is set, the meter will automatically shut down after 15 minutes if no operation; when 
“Off” is set, the auto. power off function will be turned off. Long press 

  to manually shut down 

the meter.

 

5) 

pH Calibration Reminder (P07) 

– set X hours (H) Or X days (D) in ZenTest mobile app – settings 

– Parameter – pH – Calibration Reminder. On the meter, you can only check the values that’s been 
set up on ZenTest App. For example, if 3 days is set up, the Er6 icon (see Figure-4) will appear in 
the lower right corner of the LCD screen in 3 days to remind you to perform calibration, also in the 
ZenTest App there will be a pop-up reminder. After calibration is finished or the reminder setting is 
cancelled in the ZenTest App, the Er6 icon will disappear.

 

Symbol 

Parameter Setting Contents 

Content 

Factory Default 

P1 

Temperature Unit 

˚C –

 

˚F

 

˚F

 

P2 

Select automatic lock 

5-20 seconds 

 Off 

Off 

P3 

Automatic Backlight Off 

1-8 minutes

 

 Off 

P4 

Automatic Power Off 

10-20 minutes 

 Off 

10 

P5 

pH Buffer Series Selection 

USA 

 NIST 

USA 

P6 

pH Resolution 

0.1 

 0.01 

0.01 

P7 

pH Calibration Reminder 

H-hours D-Days (set up in 

ZenTest App) 

P8 

pH back to factory default 

No

 

 Yes 

No 

Summary of Contents for AI3711

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...Calibration 5 7 pH Measurement 7 8 ORP Measurement 8 9 Probe Cleaning 8 10 Probe Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Tr...

Page 3: ...connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to sw...

Page 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Page 5: ...calibration bottles to about half volume of the bottle Long press to enter calibration mode screen turns green Insert the probe into pH 7 00 buffer solution make a quick stir and hold still When the r...

Page 6: ...H first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 10 times as long as they are not contaminated and the bottles are capped when not in use pH 10 0...

Page 7: ...ibration again 3 The electrode is aged used for over a year and has a much slower response A replacement is needed pH electrode zero electric potential out of range 60mV or 60mV 1 Check whether pH buf...

Page 8: ...d to be able to measure ORP 2 Power on the PC60 Z tester unscrew the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14...

Page 9: ...rt press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut down af...

Page 10: ...ature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV Accuracy 0 2 F S Temperature Range 0 to 50 C 32 122 F Accuracy 0 5 C Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6...

Page 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Page 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...