  For food pH measurement (e.g. sushi, cheese, meat, fruit, etc.), it should always be a sampling 
test. That means test samples should no longer be edible.

 

 

  How to take pH measurements   

Short press 

  to turn on the tester. Rinse the probe in distilled water, shake off excess water. Insert 

the probe in your sample solution, make a quick stir and  hold still. Record the reading when it is 
stabilized (

appears and stays on screen).

 

 

  Pure Water pH Measurement 

When testing pure water like tap water, drinking water, RO water and distilled water, it will take longer 
for the readings to get fully stabilized (typically 1-5 minutes). Please be patient. If still not working, add 
Apera 3M KCL (AI1107) to your pure water at the ratio of 1:1000 (e.g. 1 ml KCL to 1000 ml water) to 
accelerate stabilization while minimizing pH change. If the accuracy is not meeting your requirement, 
please contact us at [email protected] to find the specialized meter designed for pure water pH test.

 

8.  ORP Measurement 

1) 

ORP stands for Oxidation-Reduction Potential, measured in mV. It’s also called redox. ORP

 

is a 

measure of the cleanliness of water & its ability to break down contaminants. A separate ORP 
probe (ORP60-DA) needs to be installed to be able to measure ORP. 

 

2) 

Power on the PC60-Z tester, unscrew the original probe, and install the ORP60-DA probe, then 
the tester will automatically switch to ORP measurement mode (Refer to Section 14 for how to 
replace a probe).

 

3) 

Rinse the probe in distilled water and dry it. Dip the probe in sample solution, shake for a few 
seconds, and allow it to stand still. Get the ORP readings after 

appears and stays on screen.

 

9.  Probe Cleaning 

1) 

The tester is only as accurate as the probe is clean. Always thoroughly rinse off the probe before 
and after each measurement with pure water in a container or with a wash bottle. 

 

2) 

For  tough  contaminants,  detach  the  sensor  shield,  soak  the  probe  in Apera  probe  cleaning 
solution  (AI1166)  or  detergent  water  for  30  minutes.  Then  use  a  soft  brush  to  remove  the 
contaminants. Afterwards, soak the probe in 3M KCL soaking solution for at least 1 hour. Rinse it 
off, then re-calibrate the tester before using again.

 

10. Probe Storage 

1)  Always store the spear probe in the 3M KCL soaking solution when not in use. Never let it dry 

out. Replace the soaking solution if it gets contaminated.

 

2)  NEVER store the probe in pure water like tap, RO, distilled, or deionized water as they could 

damage the pH probe. If this happens, immediately soak the pH probe Apera 3M KCL soaking 

solution overnight, then re-calibrate it before using. Pure water is only for rinsing the probe.

 

Summary of Contents for AI3711

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...Calibration 5 7 pH Measurement 7 8 ORP Measurement 8 9 Probe Cleaning 8 10 Probe Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Tr...

Page 3: ...connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to sw...

Page 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Page 5: ...calibration bottles to about half volume of the bottle Long press to enter calibration mode screen turns green Insert the probe into pH 7 00 buffer solution make a quick stir and hold still When the r...

Page 6: ...H first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 10 times as long as they are not contaminated and the bottles are capped when not in use pH 10 0...

Page 7: ...ibration again 3 The electrode is aged used for over a year and has a much slower response A replacement is needed pH electrode zero electric potential out of range 60mV or 60mV 1 Check whether pH buf...

Page 8: ...d to be able to measure ORP 2 Power on the PC60 Z tester unscrew the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14...

Page 9: ...rt press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut down af...

Page 10: ...ature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV Accuracy 0 2 F S Temperature Range 0 to 50 C 32 122 F Accuracy 0 5 C Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6...

Page 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Page 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...

Reviews: