- 12 -

 

Display similar 

readings in any 

solutions or 

always display 

7.0 pH 

Broken probe 

If you don’t find any visible damage of the probe, 

contact us for warranty fulfillment; If there is visible 

damage, replace the probe. 

Reading keeps 

jumping 

Probe is not fully 

submerged in the solution 

Make sure the probe is fully immersed in the 

solution at least 1 inch deep. 

Air bubbles around the 

sensor   

Make a quick stir in the solution to remove air 

bubbles. 

Probe is not properly 

connected, or the pin 

connector is broken. 

Check the 

probe’s connector, make sure it’s not 

broken and is correctly connected. Align the probe 

and instrument correctly before plugging in. Never 

force it. Ensure that the probe connector is not 

exposed to air too long. 

Calibration is 

successful, but 

reading is not 

accurate 

Aged probe 

Replace the probe. 

Air bubbles around the 

sensor 

Make a quick stir in the solution to remove air 

bubbles. 

Clogged junction 

Clean the probe with cleaning solution, then soak it 

in 3M KCL soaking solution overnight. 

Comparison with other 

testers, test strips, or drop 

tests 

To compare with other testers, make sure to 

perform a 2-point calibration for all testers in the 

same standard, then test a 3rd point. Whichever 

gives more accurate reading is the more accurate 

one. Test strips or drop tests’ accuracy is not 

comparable to pH meters ‘.

 

16. Warranty 

We warrant this instrument to be free from defects in material and workmanship and agree to repair 
or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or 
damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a period 
of TWO YEARS (SIX MONTHS for the probe) from the delivery. This limited warranty does NOT cover 
any  damages  due  to:  accidental  damage,  unauthorized  repair,  normal  wear  and  tear,  or  external 
causes such as accidents, abuse, or other actions or events beyond our reasonable control.

 

 
 
 
 
 
 
 
 
 
 
 

APERA INSTRUMENTS (Europe) GmbH

 

Wilhelm-Muthmann-Straße 18, 42329 Wuppertal, Germany

 

Contact: 

[email protected] |

 

www.aperainst.de

 |Tel. +49 202 51988998

 

Summary of Contents for AI3711

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...Calibration 5 7 pH Measurement 7 8 ORP Measurement 8 9 Probe Cleaning 8 10 Probe Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Tr...

Page 3: ...connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to sw...

Page 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Page 5: ...calibration bottles to about half volume of the bottle Long press to enter calibration mode screen turns green Insert the probe into pH 7 00 buffer solution make a quick stir and hold still When the r...

Page 6: ...H first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 10 times as long as they are not contaminated and the bottles are capped when not in use pH 10 0...

Page 7: ...ibration again 3 The electrode is aged used for over a year and has a much slower response A replacement is needed pH electrode zero electric potential out of range 60mV or 60mV 1 Check whether pH buf...

Page 8: ...d to be able to measure ORP 2 Power on the PC60 Z tester unscrew the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14...

Page 9: ...rt press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut down af...

Page 10: ...ature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV Accuracy 0 2 F S Temperature Range 0 to 50 C 32 122 F Accuracy 0 5 C Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6...

Page 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Page 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...

Reviews: