pH Back to Factory Default (P08) 

– Select “Yes” to recover instrument calibration to theoretical 

value. This function can be used when instrument does not work well in calibration or measurement. 
Calibrate and measure again after setting the instrument back to factory default. 

 

12. Technical Specifications 

13. Icons and Functions 

 

 

 
 
 

pH 

Range 

-2.00 to 16.00 pH 

Resolution 

0.01 pH 

Accuracy 

±0.01 pH ±1 digit 

Calibration Points 

1 to 3 points 

Auto. Temperature Compensation 

 

50˚C (32 –

 

122˚F)

 

ORP (mV) 

Range 

-1000 mV to 1000 mV 

Accuracy 

±0.2% F.S 

Temperature 

Range 

0 to 50˚C (32

-

122˚F)

 

Accuracy 

±0.5˚C

 

Calibrated points

 

 

Self-Diagnosis 

Symbol

 

Er1, Er2, Er3, Er4, Er5, Er6

 

Stable reading indicator

 

 

Waterproof Rating 

IP67, floats on water 

Reading Lock

 

HOLD

 

Power 

DC3V, AAA batteries*4 

Bluetooth Signal

 

 

Battery Life 

200 Hours 

Low power reminder

 

         

Backlight 

White: Measurement.   

Green: Calibration; Red: Alarm 

Auto. Power Off 

Automatically power off if no operation for 10 minutes

 

Dimension/Weight 

Instrument

40×40×178mm/133g; case

255×210×50mm/680g; 

Graph-3

 

LCD Display

 

Graph-4

 

pH calibration reminder

 

 

Graph-5

 

pH alarm triggered

 

 

Summary of Contents for AI3711

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...Calibration 5 7 pH Measurement 7 8 ORP Measurement 8 9 Probe Cleaning 8 10 Probe Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Tr...

Page 3: ...connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to sw...

Page 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Page 5: ...calibration bottles to about half volume of the bottle Long press to enter calibration mode screen turns green Insert the probe into pH 7 00 buffer solution make a quick stir and hold still When the r...

Page 6: ...H first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 10 times as long as they are not contaminated and the bottles are capped when not in use pH 10 0...

Page 7: ...ibration again 3 The electrode is aged used for over a year and has a much slower response A replacement is needed pH electrode zero electric potential out of range 60mV or 60mV 1 Check whether pH buf...

Page 8: ...d to be able to measure ORP 2 Power on the PC60 Z tester unscrew the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14...

Page 9: ...rt press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut down af...

Page 10: ...ature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV Accuracy 0 2 F S Temperature Range 0 to 50 C 32 122 F Accuracy 0 5 C Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6...

Page 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Page 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...

Reviews: