Error Messages and Warnings
Error message
Cause
Out of Specification
The measured sample temperature is outside the
specifications.
Out of Range
The measured value is outside the specified range.
Possible reasons:
•
The temperature is too high or too low.
•
The density is too high or too low.
•
The measured value is invalid.
Temperature Range
The temperature of the reference liquid for
readjustment is outside the temperature
specifications (+15 to +25 °C).
Density Range
The density of deionized water at 20 °C measured
during readjustment is outside the allowed limits
(tolerance 0.01 g/cm
3
).
Criterion
The density of deionized water at 20 °C measured
during readjustment deviates from factory
adjustment by more than 0.01 g/cm
3
.
Low Battery
The battery is almost empty.
Battery Empty
The battery is empty.
Error 01
The temperature is invalid.
Error 02
The period is invalid.
Error 03
DMA 35 has no adjustment data.
Error 04
Calculation error.
NOTICE
If the error messages "Error 01" to "Error 04" occur, send in your
instrument for service.
Summary of Contents for 84138
Page 2: ......
Page 3: ...C96IB01 3 EN DMA 35 Portable Density Meter From Firmware Version V1 79 Instruction Manual ...
Page 72: ...72 C96IB01 EN Appendix C CE Declarations of Conformity ...
Page 73: ...C96IB01 73 EN ...
Page 74: ...74 C96IB01 EN Appendix D EC Type Examination Certificate ...
Page 75: ...C96IB01 75 EN ...
Page 76: ...76 C96IB01 EN Appendix E RFID Module Test Reports ...
Page 77: ...C96IB01 77 EN ...
Page 81: ...C96IB01 81 DE DMA 35 Tragbares Dichtemessgerät Ab Firmwareversion V1 79 Betriebsanleitung ...
Page 153: ...C96IB01 153 DE Anhang C CE Konformitätserklärung ...
Page 154: ...154 C96IB01 DE ...
Page 155: ...C96IB01 155 DE Anhang D EG Baumusterprüfbescheinung ...
Page 156: ...156 C96IB01 DE ...
Page 157: ...C96IB01 157 DE Anhang E RFID Modul Testreports ...
Page 158: ...158 C96IB01 DE ...