12.3 Custom Adjustment
You can also use any other reference liquid with a known density at 20 °C and known
temperature coefficient for your readjustment.
We recommend using the customer adjustment e.g. if you measure only liquids in a
specific density range (e.g. high density) and want to yield more accurate results in
this range.
1.
Clean the measuring cell as described in Chapter 10.1 and rinse the measuring
cell until no more solvent residues are present.
2.
Press <Menu> and select
"Adjustment > Custom Adjustment"
.
3.
Fill the measuring cell with reference liquid at approx. 20 °C (15 to 25 °C)
without bubbles.
4.
Confirm with <OK> that the measuring cell is filled with reference liquid.
You are asked to enter the density of your reference liquid at 20 °C.
5.
Using the arrow keys, select the first number and press <OK>.
6.
Repeat the procedure for the other numbers and press <Save>.
7.
Enter the temperature coefficient of your reference liquid and press <Save>.
The density of the filled reference liquid is now measured and compared with the
entered density for the reference liquid at 20 °C. The display shows the
calculated deviation (Delta) in g/cm
3
.
8.
Press <OK> to activate the new adjustment.
12.4 Defining an Offset
You can define an offset for your measurement. This is automatically added to each
measured value. The offset value always refers to the current measuring unit and is
deleted when you change to another measuring unit. If you have selected a method,
the set offset is saved with the method.
1.
Press <Menu> and select
"Adjustment > Custom Offset"
.
2.
Using the arrow keys, select the first number of your offset value and press
<OK>.
3.
Repeat the procedure for all other numbers and press <Save>.
The set offset is now shown on the main screen.
TIP
The offset value saved with a method can also be changed later on.
Summary of Contents for 84138
Page 2: ......
Page 3: ...C96IB01 3 EN DMA 35 Portable Density Meter From Firmware Version V1 79 Instruction Manual ...
Page 72: ...72 C96IB01 EN Appendix C CE Declarations of Conformity ...
Page 73: ...C96IB01 73 EN ...
Page 74: ...74 C96IB01 EN Appendix D EC Type Examination Certificate ...
Page 75: ...C96IB01 75 EN ...
Page 76: ...76 C96IB01 EN Appendix E RFID Module Test Reports ...
Page 77: ...C96IB01 77 EN ...
Page 81: ...C96IB01 81 DE DMA 35 Tragbares Dichtemessgerät Ab Firmwareversion V1 79 Betriebsanleitung ...
Page 153: ...C96IB01 153 DE Anhang C CE Konformitätserklärung ...
Page 154: ...154 C96IB01 DE ...
Page 155: ...C96IB01 155 DE Anhang D EG Baumusterprüfbescheinung ...
Page 156: ...156 C96IB01 DE ...
Page 157: ...C96IB01 157 DE Anhang E RFID Modul Testreports ...
Page 158: ...158 C96IB01 DE ...