C96IB01
29
EN
9
Performing a Measurement
General instructions for measuring
•
Make sure that the measuring cell is filled free from gas bubbles. Possible
reasons for gas bubbles in the measuring cell are:
•
Gas bubbles in the sample
•
Leaky connection of the filling tube, the pump or the screw plug
•
Make sure that the measuring cell is fully filled.
•
Make sure that the sample temperature does not deviate too strongly from the
ambient temperature.
When measuring samples with a temperature which greatly differs from the
ambient temperature and a measurement mode with a temperature stability
criterion is selected, the measurement takes longer because the sample
temperature first has to adjust to ambient temperature.
If the sample measured is outside the temperature specifications (0 to +40 ° C),
the display shows the warning sign until the sample has reached a measurable
temperature. If the measured value is nevertheless saved, it is shown with an
exclamation mark.
WARNING
Handling samples with temperatures of more than 70 °C bears the
danger of heavy burns. Make sure you wear protective clothes or
ensure alternative protection from burns when handling high
temperature samples.
NOTICE
Before performing a measurement, make sure that the wetted parts
are resistant to the sample (see Chapter 15).
NOTICE
Sample containing dissolved CO
2
will cause bubbles within the
measuring cell with the effect of invalid measurement results. Make
sure your sample is degassed carefully.
Degassing a sample can be done by:
•
boiling the sample for several minutes,
•
stirring the sample for 5 to 15 minutes until no bubbling occurs
any more or
•
putting the sample for approximately 5 to 10 minutes into an
ultrasonic bath.
Summary of Contents for 84138
Page 2: ......
Page 3: ...C96IB01 3 EN DMA 35 Portable Density Meter From Firmware Version V1 79 Instruction Manual ...
Page 72: ...72 C96IB01 EN Appendix C CE Declarations of Conformity ...
Page 73: ...C96IB01 73 EN ...
Page 74: ...74 C96IB01 EN Appendix D EC Type Examination Certificate ...
Page 75: ...C96IB01 75 EN ...
Page 76: ...76 C96IB01 EN Appendix E RFID Module Test Reports ...
Page 77: ...C96IB01 77 EN ...
Page 81: ...C96IB01 81 DE DMA 35 Tragbares Dichtemessgerät Ab Firmwareversion V1 79 Betriebsanleitung ...
Page 153: ...C96IB01 153 DE Anhang C CE Konformitätserklärung ...
Page 154: ...154 C96IB01 DE ...
Page 155: ...C96IB01 155 DE Anhang D EG Baumusterprüfbescheinung ...
Page 156: ...156 C96IB01 DE ...
Page 157: ...C96IB01 157 DE Anhang E RFID Modul Testreports ...
Page 158: ...158 C96IB01 DE ...