34
C96IB01
EN
9.2
Selecting Measuring Methods
Application of the method selection
You can define and store measuring methods in your DMA 35, providing a great
contribution to a more efficient organization of your measuring process, especially, if
you want to measure many different substances according to different measuring
units.
A stored measuring method may contain the following kinds of information:
•
Measuring unit
•
Offset value
•
Temperature coefficient
•
Number of the RFID tag
3
You can define and name up to 20 different measuring methods. To see how many of
the 20 storage places are still available, go to the menu
"Device Information"
(see
If you own several DMA 35 instruments, you can import the same method list to all
instruments.
9.2.1 Entering a New Method
1.
Select the preferred measuring parameter (see Chapter 9.1) and, if necessary,
the custom offset (see Chapter 12.4).
2.
Press <Menu> and select
"Methods > Enter new Method"
.
3.
Enter the first character of the measuring method's name using the arrow keys
and press <OK>.
4.
Repeat this procedure for the other characters.
5.
Navigate to the tick symbol using the arrow keys and press <Save>.
3
Only for DMA 35 versions with RFID interface
TIP
If you choose the tick symbol without first choosing the characters, the
method will be saved as <empty>. In this case, while being in the main
screen only an empty bar indicates that a method has been selected.
TIP
If you keep an arrow key pressed, the cursor automatically stops at the tick
symbol.
Summary of Contents for 84138
Page 2: ......
Page 3: ...C96IB01 3 EN DMA 35 Portable Density Meter From Firmware Version V1 79 Instruction Manual ...
Page 72: ...72 C96IB01 EN Appendix C CE Declarations of Conformity ...
Page 73: ...C96IB01 73 EN ...
Page 74: ...74 C96IB01 EN Appendix D EC Type Examination Certificate ...
Page 75: ...C96IB01 75 EN ...
Page 76: ...76 C96IB01 EN Appendix E RFID Module Test Reports ...
Page 77: ...C96IB01 77 EN ...
Page 81: ...C96IB01 81 DE DMA 35 Tragbares Dichtemessgerät Ab Firmwareversion V1 79 Betriebsanleitung ...
Page 153: ...C96IB01 153 DE Anhang C CE Konformitätserklärung ...
Page 154: ...154 C96IB01 DE ...
Page 155: ...C96IB01 155 DE Anhang D EG Baumusterprüfbescheinung ...
Page 156: ...156 C96IB01 DE ...
Page 157: ...C96IB01 157 DE Anhang E RFID Modul Testreports ...
Page 158: ...158 C96IB01 DE ...