Cleaning and Storing the Instrument
10.1 Cleaning the Measuring Cell
Clean the measuring cell regularly before and after each measurement series with a
suitable solvent to ensure the long-term accuracy of your results. If the measuring cell
is not sufficiently cleaned, residue may form in the measuring cell and lead to
inaccurate measurement results. Depending on the application, also cleaning
between measurements may be required.
1.
Place the filling tube in a vessel which is suitable for sample disposal.
2.
Empty the measuring cell by pressing the pump lever.
3.
Fill the measuring cell with a suitable solvent.
4.
Pump the solvent through the whole measuring system several times.
5.
Empty the measuring cell by pressing the pump lever.
10.2 Cleaning the Filling Pump
Clean the filling pump regularly, depending on your application. If you measure
aggressive samples, e.g. battery acid, clean the filling pump more often.
1.
Place the filling tube in a vessel which is suitable for sample disposal.
2.
Empty the measuring cell by pressing the pump lever.
3.
Open the pump lock and remove the pump.
4.
Rinse the pump under running tap water while moving the pump piston up and
down.
5.
Dry the pump with a lint-free cloth.
6.
Insert the pump and close the pump lock.
7.
Empty the measuring cell by pressing the pump lever.
10.3 Cleaning the Pump Housing
DMA 35 is completely sealed against the outside and the pump. Therefore, you can
rinse the pump housing with running tap water if the housing is dirty.
1.
Place the filling tube in a vessel which is suitable for sample disposal.
NOTICE
Make sure the solvent you use for cleaning is suitable. The DMA 35
pump cylinder is not resistant to aggressive cleaning liquids such as
acetone or methyl ethyl ketone (MEK).
Summary of Contents for 84138
Page 2: ......
Page 3: ...C96IB01 3 EN DMA 35 Portable Density Meter From Firmware Version V1 79 Instruction Manual ...
Page 72: ...72 C96IB01 EN Appendix C CE Declarations of Conformity ...
Page 73: ...C96IB01 73 EN ...
Page 74: ...74 C96IB01 EN Appendix D EC Type Examination Certificate ...
Page 75: ...C96IB01 75 EN ...
Page 76: ...76 C96IB01 EN Appendix E RFID Module Test Reports ...
Page 77: ...C96IB01 77 EN ...
Page 81: ...C96IB01 81 DE DMA 35 Tragbares Dichtemessgerät Ab Firmwareversion V1 79 Betriebsanleitung ...
Page 153: ...C96IB01 153 DE Anhang C CE Konformitätserklärung ...
Page 154: ...154 C96IB01 DE ...
Page 155: ...C96IB01 155 DE Anhang D EG Baumusterprüfbescheinung ...
Page 156: ...156 C96IB01 DE ...
Page 157: ...C96IB01 157 DE Anhang E RFID Modul Testreports ...
Page 158: ...158 C96IB01 DE ...