24
24
EN
R7
- OPERATING INSTRUCTION
5.1.1.2.1. Hygiene system settings
NOTE:
menu available only if at least a hygiene system is present.
From the GENERAL SETTINGS menu touch the icon button
to
access the HYGIENE SYSTEM SETTINGS submenu containing the
following icon buttons:
BIOSTER /S disinfection cycle setting
(only if the BIOSTER /S system is present )
Flushing CYCLE SETTING
(only if the FLUSHING system is present)
5.1.1.2.1.1. BIOSTER /S disinfection cycle setting
This setting is shared by all users.
From the HYGIENE SYSTEM SETTINGS menu carry out the following
operations:
• Touch the icon button
to access the BIOSTER/S DISINFECTION
CYCLE SETTING submenu.
• Withdraw the instruments to be treated (the corresponding icon will appear
on the display):
S1: syringe on instrument board.
A: instrument in position A
B: instrument in position B
C: instrument in position C
D: instrument in position D
F: instrument on assistant's board.
BC: water to cup duct. ( always operative )
The syringe on the assistant's board cannot be disinfected using
the BIOSTER S cycle.
•
To start the disinfection cycle, touch the icon button
(see Paragraph
7.4).
NOTE: the disinfectant contact time is preset and not modifiable
(600 seconds).
NOTE: the BIOSTER S cycle can only be activated if the SHS/S
system is on (see paragraph 7.2).
Summary of Contents for SMART TOUCH R7
Page 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Page 2: ......
Page 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Page 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Page 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Page 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Page 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Page 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Page 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Page 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Page 90: ......