3.1.
R7 line operating unit conversion
To convert the operating unit from the right-handed operator
version to the left-handed operator version, proceed as follows:
• First of all, set up the operatory unit for conversion by moving
the backrest completely up, raising the dental chair at least more
than half of its vertical stroke and placing the instrument board’s
arm and body in the middle in relation to the dental chair seat
so that it is out of the way.
WARNING!
At this point, shut off the operatory unit so that it cannot
accidentally be turned on, creating hazardous situations.
• Take off the cover (
a
) that protects the hydrogroup’s short support
arm and blocks movement.
• Lift the element (
k
) locking the short arm in place.
NOTE
: slightly move arms on their pivot points so as to
make element release easier.
• Turn the assistant’s board’s arms until they are compact in
relation to the hydrogroup.
• Start turning the hydrogroup and relative arm.
WARNING!
During this operation, avoid extending the arms of the
operating lamp, if any, to the outside so as to prevent any
excessive imbalance of the equipment.
• Once the hydrogroup is aligned with the dental chair, turn the
assistant’s board’s arms (
b
) bringing them into the new work
position.
WARNING!
Be extremely careful not to pinch the suction tubes and
instrument cords while performing this operation.
• Finish turning the hydrogroup to the left bringing it to the normal
work position for left-handed operators.
NOTE
: make sure that element (
k
) is correctly locking short
arm in the new position.
• Turn the bowl to the new home position by hand.
• Place the pin (
c
), that stops rotation of the instrument board’s
arm, in the opposite hole to correctly set the arm’s work area.
WARNING!
If the stop pin is not used, the arm may bang against the
hydrogroup damaging it during regular operation.
• Put the cover (
a
) back on making sure the cover’s pins are
properly placed in the seats in the short arm thereby blocking
the hydrogroup.
• The operatory unit is now ready to be used by left-handed ope-
rators.
WARNING!
Before attempting to turn on the operatory unit, verify all
the equipment is in the correct work condition.
Summary of Contents for SMART TOUCH R7
Page 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Page 2: ......
Page 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Page 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Page 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Page 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Page 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Page 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Page 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Page 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Page 90: ......