51
Cycle
LED
Total
time
Ø8 mm
Total
energy
standard
1
20"
1.000 mW/cm
2
20.000 mJ
fast
2
15"
1.600 mW/cm
2
24.000 mJ
strong
3
20"
1.800 mW/cm
2
36.000 mJ
bonding
S+1
15"
ramp cycle
11.250 mJ
rapid rest.
S+2
15"
ramp cycle
20.250 mJ
long rest.
S+3
20"
ramp cycle
26.250 mJ
51
EN
R7
- OPERATING INSTRUCTION
5.7.
T LED curing light
(not available for the American and Canadian markets)
Technical specifications.
Supply voltage: 24-36 VDC
Max. power absorbed: 6 VA
Light source: 1 5W LED
Wavelength: 430-490 nm
Acoustic signals: at cycle start, every 5 seconds, and at cycle end
Type of operation : intermittent (3 consecutive cycles - 60 sec. rest)
Programs: 6 (preset)
General description of the light
a) Light handpiece
b) Rotary end section
c) Fiber optic
d) Eye protection
e) Power cord
f) Start button
NOTE:
The curing light can be used in different configurations (wand,
gun or any intermediate position) to aid the user.
NOTE:
The curing light is delivered in its original packing which
should be kept for future shipment.
Description of the control pad
[ 1 ] LED 1 ( STANDARD cycle ) :
Emission of 1000 mW/cm
2
for 20 seconds (this cycle is set as default
at the time of sale).
[ 2 ] LED 2 ( FAST cycle ) :
Emission of 1600 mW/cm
2
for 15 seconds.
[ 3 ] LED 3 ( STRONG cycle ) :
Emission of 1800 mW/cm
2
for 20 seconds.
[ 4 ] LED S :
When LED S is on, you access ramp cycle mode and at the same time
the LEDs B, R and L next to it come on:
[ LED S + LED 1 ] ramp cycle B ( BONDING ) :
Ramp cycle with emission of 500 mW/cm
2
for 5 seconds, ramp from
500 to 1000 mW/cm
2
for 5 seconds and 1000 mW/cm
2
for 5 seconds
for a total of 15 seconds.
[ LED S + LED 2 ] ramp cycle R ( RAPID RESTORATION ) :
Ramp cycle with emission of 500 mW/cm
2
for 5 seconds, ramp from
500 to 2200 mW/cm
2
for 5 seconds and 2200 mW/cm2 for 5 seconds
for a total of 15 seconds.
[ LED S + LED 3 ] ramp cycle L ( LONG RESTORATION ) :
Ramp cycle with emission of 500 mW/cm
2
for 5 seconds, ramp from
500 to 1800 mW/cm
2
for 5 seconds and 1800 mW/cm
2
for 10 seconds
for a total of 20 seconds.
[ 5 ] Malfunction signalling LED:
This red LED comes on only if there is a malfunction.
[ 6 ] START button:
Pressing the START button starts the cycle selected at that moment
(the cycle indication LED will come on).
If it is pressed again at any time during the cycle, light beam emission
will immediately be interrupted.
[ 7 ] MODE button:
This button is used to select the cycle to be run. It allows changing
from the cycle you are in at that moment to the immediately following
cycle.
The first three cycles (1, 2 and 3) are at constant power and the LEDs
come on individually.
When LED S is on, you access ramp cycle mode and at the same time
the LEDs B, R and L next to it come on.
Once the LED of the cycle you intend to use has come on, the lamp
is ready for use. Pressing the START button, light beam emission is
activated according to the cycle selected .
NOTE
: the cycle can be selected and the button is operative
only when the curing light is not emitting any light. If the button is
accidentally pressed while light is being emitted, nothing will happen.
Operation.
WARNING!
The instrument is supplied non-sterile.
Before use, disinfect the lamp grip. The optical fibre and the eye pro
-
tection can be sterilized in a steam autoclave at 135°C.
• Put the fiber optic (
c
) in its housing until it clicks.
• Attach the curing light handpiece to the end of its power cord and tighten
the ring (
e
).
Summary of Contents for SMART TOUCH R7
Page 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Page 2: ......
Page 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Page 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Page 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Page 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Page 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Page 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Page 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Page 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Page 90: ......