19
19
EN
R7
- OPERATING INSTRUCTION
Inverting the console unit position for left-handed operators
WARNING!
BEFORE CARRYING OUT THIS OPERATION, TURN OFF THE DENTAL
UNIT. DO NOT REMOVE THE CONSOLE FROM THE DENTIST’S MO-
DULE IF THE DENTAL UNIT IS ON.
To invert the position of the console unit on the dentist’s module,
operate as follows:
• Remove the console unit after unscrewing the fastening ring nut (
g
) by
turning it anticlockwise.
• Remove the protective snap-on cap (
s
) of the quick-coupling on the left-
hand side and fit it on the coupling on the right-hand side.
• Turn the console unit brace 180°.
• Fit the console unit to the quick-coupling on the left-hand side.
To correctly position the console unit, push the brace fully into place in the
support and at the same time turn the ring nut by about 1/3 turn until it locks
without forcing tightening.
NOTE:
to prevent the dentist’s module from slipping out on the
opposite side during this operation, it is advisable to beforehand turn it
about 90° with respect to its support arm (see figure).
• At this point, you can turn the dental unit on again.
WARNING!
During the console cleaning operations do not exert excessive pressure
on the pushbutton panel to prevent damaging stress on the connection.
Cleaning the dentist’s instrument board.
Clean the dentist’s instrument board using a suitable product (see para-
graph 1.4).
NOTE for CONTINENTAL version dentist’s instrument boards:
the instrument holder (
x
) can be removed to facilitate the cleaning ope-
rations; to remove it, simply pull it out of its seat as it is only secured with
magnets.
The silicone instrument holder ( u ) can also be sterilized in an autoclave
at 121°C (see paragraph 1.5.).
Removable instrument cords
All the instruments have removable cords to ease cleaning.
NOTE for CONTINENTAL version dentist’s instrument boards:
to remove the tubings you first need to remove the instrument holder and
then unscrew the relative plastic fastening ring nuts.
NOTE for INTERNATIONAL version dentist’s instrument bo-
ards:
to remove the tubings, unscrew the relative plastic fastening ring
nuts below the dentist’s instrument board.
WARNING!
• Shut off the operatory unit before attempting to take off the cords.
• After shutting off the operatory unit, empty the syringe’s ducts by
pressing and holding down the relative air and water buttons directly
on the bowl until water spray is no longer present.
• The cords of the TURBINE, MICROMOTOR and SCALER contain
water, therefore hold the end of the cord on the handpiece side over
the bowl when removing the cord.
• When putting a cord back on, make certain the contacts are perfectly
dry and the plastic ring nut is tight.
• Each cord may be remounted only in the position for the correspon-
ding instrument.
Clean the instrument cord using a suitable product (see Paragraph 1.4).
WARNING!
The instrument cords are NOT suitable for autoclave or cold disin-
fection.
Summary of Contents for SMART TOUCH R7
Page 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Page 2: ......
Page 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Page 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Page 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Page 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Page 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Page 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Page 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Page 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Page 90: ......