5 SOFTWARE-UPDATE
CAUTION:
The software update of the
eLockIn
can be compared with a
BIOS
update
of a PC. One should spare the same caution. A power failure or power blackout
during this procedure might cause a system crash, after which the
eLockIn
software
needs to be restored in our facilities.
When this procedure fails, the instrument
has to be shipped to us!
If all precautions against power failure are done, follow these steps:
Unpack the software update.
Create a directory “update” (case-sensitive!).in the root on a USB stick.
Copy all data into this directory.
Plug the USB-Stick in the front panel of the eLockIn20-dual and wait a few seconds.
Chose “SETUP”. Press the knob besides “UPDATE”. The update starts. The
eLockIn204/2 displays all files which are uploaded from the stick in a list.
The update is completed, if it displays “DONE”.
Switch the
eLockIn
OFF and wait for at least 5 to 10 s.
Switch the instrument ON again. The new software should run.