Result for „GET /setup/lockin.ini CR
CR“
4.5.2 DLL
BASED
ACCESS
BY
L
AB
V
IEW
eLockin.vi hierarchie
The main program eLockin.vi contains a sequence in
three steps and calls two sub-VI's: Ref_eLockin.vi and
Get_eLockin.vi.
During run, it calls a three-step sequence:
●
Step 0:
eLockin.vi starts the communication
with the lockin. The IP address is a value entry
in the diagram used as input for the library
function call „__Init“. Then, the system waits for
500 ms to let the
eLockIn
initialize.
●
Step 1:
The Sub-VI „Ref_eLockin.vi“ is called to set reference voltage and
reference frequency. The other two functions of this Sub-VI, phase and
harmonic, are not used here.
●
Step 2:
A curve diagram is drawn with 100 values, which are taken from the
Sub-VI „Get_eLockin.vi“, which uses the function call of „__GetLockInData2 for
the channel number 1. Between the single data, 100 ms delay time is given.
The example does not use all provided DLL-functions, but it shows how to implement
them in own applications.
Usage of the LabView control
To start LabView, double-click on the icon or program eLockin.vi.
The main program shows a curve diagram and two input windows for
the reference amplitude and reference frequency:
The entered values are overtaken, when the start knob
is applied.
The same knob starts the acquisition of data in the curve diagram for
one time cycle.
48 (58)
Summary of Contents for eLockIn 203
Page 52: ...7 APPENDICES 7 1 SCHEMATIC DIAGRAM OF THE SIGNAL PATHS 52 58...
Page 53: ...7 2 DIAGRAM OF THE LOCKIN AMPLIFICATION PATHS Diagram of the preamplifier switches 53 58...
Page 54: ...7 3 FRONTPANEL DIMENSIONS 54 58...
Page 55: ...7 4 INSTRUMENT SPECIFIC TEST SHEETS insert with 10 pages of data 55 58...
Page 56: ...7 5 PREAMPLIFIER CONNECTOR PINOUT 56 58...
Page 58: ...7 7 TRACEABILITY CHART 58 58...