8.7. Weighing Sequence Functions
For the function setting procedure, refer to “19. GENERAL FUNCTION SETTING.”
Sq F-
Item
Setting value
Default
01 Comparison
target
1: Internal value
2: Displayed weight value
1
02
Automatic accumulation condition
0: Disable
1: OK only
2: Regardless of the judgment result
0
03
Automatic free fall correction
0: Disable
1: Moving average of last four weighings
2: Fuzzy
0
04
Automatic free fall correction range
0 to 9999999
0
05
Minimum small flow time for the
automatic free fall correction
0.0 to 30.0 s
3.0
06
OK / Over / Under output timing
1: Always
2: In synchronization with weighing end
2
07
Wait for the weight value to be stable
before the judgment
0: Disable
1: Enable
1
08
Maximum number of correction flow
0: Disable correction flow
1 to 300
0
09
Weighing start input delay time
0.0 to 30.0 s
0.0
10
Large flow comparison disable time
0.0 to 30.0 s
0.0
11
Medium flow comparison disable time
0.0 to 30.0 s
0.0
12
Small flow comparison disable time
0.0 to 30.0 s
0.0
13
Judging delay time
0.1 to 30.0 s
0.1
14
Weighing end output time
0.0: Until the next weighing start
0.1 to 30.0 s
0.0
15
Flow timeout time
0: None
1: 1 to 86400 s
0
16
Correction flow ON time
0.01 to 3.00 s
0.10
17
Correction flow OFF time
0.1 to 30.0 s
0.1
18
Discharge start input delay time
0.0 to 30.0 s
0.0
19
Discharge end delay time
0.1 to 30.0 s
0.1
20
Discharge timeout time
0: None
1: 1 to 86400 s
0
21
Add the final weight to the near-zero
setpoint value
0: Disable
1: Enable
0