AD-4401A Page
24
●
Weighing sequence errors
When a sequence error is detected, the detected sequence number is displayed in the sub-display. (FncF-04)
Sequence No.
Description
00
Waits for the weighing start input.
01
Automatic tare
02
Weighing start input delay time (1 s) after automatic tare
03
Starts the weighing start input delay.
04
Compares the weight value with the setpoint.
05
Waits for the weighing start input delay time to pass.
06
Turns the large flow/medium flow/small flow outputs ON.
07
Waits for the large flow comparison disable time to pass.
08
Waits for the weight value to exceed the optional preliminary value.
09
Turns the large flow output OFF and waits for the medium flow comparison
disable time to pass.
10
Waits for the weight value to exceed the preliminary value.
11
Turns the medium flow output OFF and waits for the small flow comparison
disable time to pass.
12
Waits for the weight value to exceed the free fall value.
13
Waits for the judging delay time to pass.
14
Waits for the weight value to be stable.
15
Judges if the correction flow is required.
16
Judges the weight value.
17
Waits for the weight value to be near-zero.
30
Starts the correction flow.
31
Turns the small flow output ON for correction.
32
Waits for the correction flow ON time to pass and turns the small flow output
OFF.
33
Waits for the correction flow OFF time to pass.
34
Judges the correction flow value.
35
Ends the correction flow.
50
Waits for the discharge start input.
51
Waits for the discharge start input delay time to pass.
52
Turns the discharge output ON.
53
Waits for the weight value to be below the near-zero value.
54
Waits for the discharge end delay time to pass.
55
Turns the discharge output OFF.
Current sequence number
Number of the sequence in which an error is detected.