Page 11
AD-4401A
6. CALIBRATION
Calibrates the AD-4401A to convert the signal from the load cell to a mass value correctly.
6.1. Calibration Setting
Set the following calibration functions which are required for calibration using calibration weights.
●
Unit: (CALF-01)
Select the unit of the scale.
●
Decimal point position: (CALF-02)
Select the position of a decimal point.
●
Minimum division: (CALF-03)
Set the minimum division of the scale.
●
Maximum capacity: (CALF-04)
Set the maximum weighing capacity of the scale.
Once the calibration using calibration weights has been completed, the following calibration functions are
automatically set.
●
Load cell signal voltage at zero calibration: (CALF-15)
●
(Load cell signal voltage at span calibration
CALF-15)
CALF-17 / Calibration weight value at span
calibration: (CALF-16)
●
Maximum capacity at span calibration: (CALF-17)
Calibration is also performed without calibration weights by setting the load cell signal voltage corresponding
to the calibration weight value (Digital span).