6.2. Calibration Using Calibration Weights
Press the CAL key in the normal mode.
When “
Cal
" is displayed, press the ENTER key.
When “
Cal 5et
” is displayed, press the ENTER
key.
Zero calibration
Press the ENTER key when the STABLE (MD)
status is turned OFF.
Dashes are displayed in the sub-display and zero
calibration is performed.
Note:
When the Function key is pressed, the AD-
4401A proceeds to span calibration without
performing zero calibration.
Span calibration
Set the calibration weight value in the sub-display.
Press the ENTER key when the STABLE (MD)
status is turned OFF.
Dashes are displayed in the sub-display and span
calibration is performed.
Press the ON/OFF key twice to return to the normal
mode.