Error List
C Err
Description
2
The load cell signal voltage at zero calibration exceeds 2 mV/V.
Check if unnecessary load is applied to the load cell.
3
The load cell signal voltage at zero calibration is less than 0 mV/V.
Check if load cell connections of SIG
+
and SIG
-
are not reversed.
4
The calibration weight value that is set in span calibration exceeds the maximum capacity.
Set the calibration weight value to less than the maximum capacity.
5
The calibration weight value that is set in span calibration is less than the minimum division.
Set the calibration weight value to greater than the minimum division.
7
The load cell signal voltage at span calibration is less than that at zero calibration.
Check the load cell wiring.
8
The load cell signal voltage at span calibration exceeds 3.2 mV/V.
Check if unnecessary load is applied to the load cell.