4.2
List Alarm points (Displayed as List Alrm pnts)
4.2.1
List Alrm pnts (CO
2
example)
The Top-level Menu, List Alrm pnts option enables you to view read-only information about the alarm setpoints
configured for each Sensor.
NOTE:
THE List Alrm pnts OPTION PROVIDES READ-ONLY INFORMATION THAT CAN NOT BE
CHANGED. IF YOU NEED TO SET UP OR CONFIGURE THE ALARM SETPOINTS, USE THE Top-level
menu, Sensor Config, Alrm set-pnts OPTION.
1]
To display the Top-level Menu List Alrm pnts option, press and hold down the
Cycle buttons for at least 6 seconds. Then press the Cycle button once.
2]
Press Accept/Test to go to the Select Sensor screen, this displays the Sensors
and highlights Sensor 1
3]
Press Cycle to highlight other sensors
4]
Press Accept/Test to select the sensor that you want to view alarms on, then
press Cycle to cycle through the alarms
5]
Press Cancel to exit the Alarm screen and return to Select another sensor or
press Cancel to return to the Top-level Menu, List Alrm pnts