Ax60+ Multi-Gas
Service Manual
Int. Approved
Document ref: P0159-803-11
September 2019
Page 6 of 62
Copyright © 2015 Analox Ltd. All Rights Reserved.
Commercial in Confidence
1.4
Alarm terminology (CO
2
)
Alarms will display in priority order, in the order of AL4 to TWA
Display name for
users
Alarm description Identification by
software
Default
setpoint
Notes
TWA
Time weighted
average over 8
hours
Alarm 1 or A1
0.5% CO
2
Fixed alarm. At present it is not possible to
change a TWA alarm set-point.
AL1
High
alarm
Alarm 2 or A2
1.5% CO
2
This alarm must be set lower than Alarm 3
(it may be set to 0.5% to give a pre-alarm
condition).
CO2
High-high alarm
Alarm 3 or A3
3.0% CO
2
This alarm must be set higher than Alarm
2.
AL4
High-high alarm
Alarm 4 or A4
3.5% CO
2
Alarm is disabled by default.
•
The time-weighted average alarm is
displayed to users as ‘
TWA
’ and identified by the software as ‘
A1
’ or
‘
Alarm 1
’
and is announced by the Central Display only.
•
The high
alarm is displayed to users as ‘
AL1
’ and identified by the software as ‘
A2
’ or ‘
Alarm 2
’
and triggers
the Alarm strobe once every 2 seconds, sounder is disabled by default, but if enabled by the user it will also
sound once every 2 seconds.
•
The high-high
alarm is displayed to users as ‘
CO2
’ and identified by the software as ‘
A3
’ or ‘
Alarm 3
’
and
triggers the Alarm strobe/sounder once every second.
•
The fourth alarm level ‘AL4’ is disabled by default; this is identified by the software as ‘
A4
’ or ‘
Alarm 4
’. If
enabled the Alarm strobe and sounder would trigger once every second.
1.5
Alarm terminology (O
2
)
Alarms will display in priority order, in the order of AL4 to AL1
Display name for
users
Alarm description Identification by
software
Default
setpoint
Notes
AL1
Low
alarm
Alarm 1 or A1
19.5% O
2
Alarm is disabled by default.
AL2
Low
alarm
Alarm 2 or A2
19.5% O
2
This alarm must be set lower than Alarm 3.
AL3
High-high alarm
Alarm 3 or A3
23.0% O
2
This alarm must be set higher than Alarm
2.
AL4
Low-low
alarm
Alarm 4 or A4
18.0% O
2
This alarm must be set lower than Alarm 2.
•
The first low or high alarm
‘
AL1
’
AL
1
is disabled by default; this is identified by the software as ‘
A1
’ or ‘
Alarm
1
’.
If enabled the Alarm strobe and sounder would trigger once every 2 seconds.
•
The low-
level alarm is displayed to users as ‘
AL2
’ and identified by the software as ‘
A2
’ or ‘
Alarm 2
’
and
triggers the Alarm strobe once every 2 seconds, sounder is disabled by default, but if enabled by the user it
will also sound once every 2 seconds.
•
The high-high
alarm is displayed to users as ‘
AL3
’ and identified by the software as ‘
A3
’ or ‘
Alarm 3
’
and
triggers the Alarm strobe/sounder once every second.
•
The low-low alarm
is displayed to users as ‘
AL4
’ and identified by the software as ‘
A4
’ or ‘
Alarm 4
’
and
triggers the Alarm strobe/sounder once every second.