4.1.2
Show Sensor configuration (Displayed as Show Sensor config)
The Show Sensor config option enables you to view read-only information about the sensor configuration. It displays
information on Sensor alarms, sounders and strobes.
NOTE:
THE Show Sensor Config OPTION PROVIDES READ-ONLY INFORMATION THAT CAN
NOT BE CHANGED. IF YOU WANT TO SET UP OR CONFIGURE THE Ax60+, USE THE Top-level Menu,
Sensor Config OPTION.
1]
To display the Top-level Menu Show config option, press and hold down the
Cycle buttons for at least 6 seconds.
2]
Press Accept/Test, then press Cycle six times to display the Show Sensor Config?
option
3]
Press Accept/Test to enter the menu, then press Cycle to view each config
item
NOTE:
TO RETURN TO THE SYSTEM STATUS SCREEN AT ANY TIME PRESS Cancel. THE Cancel
BUTTON REDISPLAYS THE PREVIOUS MENUS, ONE AT A TIME.