6-20
P129394-176
Operator Manual
Sterilizer Operation
Table 6-4. Factory-Set
SFPP
Sterilizer Cycles and Cycle Values
Cycles
Sterilize
Temp.
Sterilize
Time
Dry
Time
Recommended Load
Validation
Standard
SFPP Cycle
270°F
(132°C)
4
Minutes
30
Minutes
Double-wrapped instrument
trays, porous and non-
porous loads. Maximum
weight per tray, 25 lb
(11.3 kg)
ST-8:2008
SFPP Cycle
270°F
(132°C)
4
Minutes
20
Minutes
*
Fabric Packs
ST-8:2008
SFPP Cycle
275°F
(135°C)
3
Minutes
30
Minutes
Double-wrapped instrument
trays, porous and
non-porous loads.
Maximum weight per tray,
25 lb (11.3 kg)
ST-8:2008
*
Dry time can be reduced to 5 minutes for a single fabric pack.
Figure 6-10. SFPP Cycle In-Cycle Screens
Cycle Selection Screen
Condition Phase
Sterilize Phase
Exhaust Phase
NOTE:
Countdown time on display is estimated; timer self-corrects estimated time at beginning
of each phase.
Press for More Cycles