4-2
P129394-176
Operator Manual
Techniques of Sterilization
Immediate Use
Prevac Cycle
270°F
(132°C)
4 Minutes
1 Minute
Unwrapped instrument tray containing
a single instrument, or multiple
unwrapped instrument trays, porous
and non-porous loads. Maximum
weight per tray: 25 lb (11.3 kg)
ST-8:2008
Gravity Cycle
250°F
(121°C)
30 Minutes 15 Minutes Fabric Packs
ST-8:2008
Gravity Cycle
270°F
(132°C)
15 Minutes
30 Minutes Double-wrapped instrument trays,
non-porous load. Maximum weight per
tray, 25 lb (11.3 kg)
ST-8:2008
Gravity Cycle
250°F
(121°C)
30 Minutes
30 Minutes Double-wrapped instrument trays,
non-porous load. Maximum weight per
tray, 25 lb (11.3 kg)
ST-8:2008
Gravity Cycle
270°F
(132°C)
25 Minutes
15 Minutes Fabric Packs
ST-8:2008
Immediate Use
Gravity Cycle
270°F
(132°C)
3 Minutes
1 Minute
Unwrapped instrument tray containing
a single instrument; or multiple
unwrapped instrument trays, non-
porous load. Maximum weight per
tray, 25 lb (11.3 kg)
ST-8:2008
Immediate Use
Gravity Cycle
270°F
(132°C)
10 Minutes
1 Minute
ST-8:2008
Liquid Cycle
250°F
(121°C)
45 Minutes
N/A
for
recommended quantities
ST-8:2008
6. Leak Test
**
N/A
N/A
N/A
None
ST-8:2008
7. Dart Test**
270°F
(132°C)
3-1/2
Minutes
1 Minute
Dart or Bowie-Dick Test Pack
ST-8:2008
8. Dart Warm-up**
270°F
(132°C)
3 Minutes.
1 Minute
None
N/A
*
Dry time can be reduced to 5 minutes for a single fabric pack.
** Not adjustable.
Table 4-1. AMSCO
®
C Series Prevacuum Sterilizer Cycles (Continued)
Cycles:
Sterilize
Temp.
Sterilize
Time
Dry
Time
Recommended Load
Refer to
Table 4 (page iii)
for
Recommended Quantities
Validation
Standard