9-21
Troubleshooting
Operator Manual
P129394-176
Alarm
Description
Screen with Operator Instructions
9.4.5 Chamber
Pressure Transducer
Failure
Occurs if chamber
pressure reading is
outside the normal
range.
Screen with
Service
Instructions
Alarm
Description
Screen with Operator Instructions
9.4.6 Chamber
Temperature Probe
Failure
Occurs if chamber
temperature reading is
outside the normal
range.
STATUS . . ALARM!
217
CHAMBER PRESSURE TRANSDUCER FAIL-
URE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
218
CHAMBER PRESSURE TRANSDUCER FAIL-
URE
-> TRANSDUCER, CP, OUTPUT VOLTAGE IS
OUTSIDE NORMAL RANGE
CAUSES AND CORRECTIONS:
1. LOOSE CONNECTION IN TRANSDUCER WIRING
-> REPAIR
2. TRANSDUCER FAILED
-> REPAIR
-> RECALIBRATE
3. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4. MAIN CONTROL FAILURE
-> CHECK CONTROL BOARD POWER STATUS LEDs
-> REPLACE CONTROL BOARD
PAPER
FEED
EXIT
STATUS . . ALARM!
211
CHAMBER TEMPERATURE PROBE FAILURE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED