8.2 Daily Maintenance
Procedures
8.2.1 Check Printer Paper
Supply
Check printer paper roll.
•
A colored warning stripe is visible when roll is near its end.
•
See
, if a new paper
roll is needed.
8.2.2 Clean Chamber Drain
Strainer
Important:
The chamber drain strainer must be cleaned at least
once a day, preferably before running the first cycle.
1.
Remove the drain strainer from the drain in the bottom of the
chamber as shown in
Figure
2.
Remove any obvious debris from the strainer. If necessary, clear
the screen in the strainer using a brush, wire or similar tool.
3.
Once it has been cleared of obvious debris, reverse flush the
strainer under running water.
4.
Replace the strainer in the chamber drain.
Figure 8-1. Remove Chamber Drain Strainer
WARNING – BURN HAZARD:
Allow sterilizer to cool to room
temperature before
performing any cleaning or
maintenance procedures.
!